This Sources Sought opportunity from Department Of Veterans Affairs was posted on May 18, 2026. The submission period has ended. Browse the details below for market research, or find similar active opportunities.
6515--Maintenance Kits for Draeger Evita V500 Ventilators
Contract Overview
Solicitation details, issuing organization, response deadlines, documents, and interested companies for this government contract opportunity.
Active Opportunities Like This One
AI Contract Overview
The Department of Veterans Affairs is conducting a market research initiative through a Request for Information (RFI) and Sources Sought Notice to identify qualified sources capable of supplying maintenance kits for Draeger Evita Infinity V500 ventilators. This RFI is explicitly not a solicitation and does not obligate the government to award a contract; responses are accepted solely for planning and acquisition decision-making purposes under FAR 15.201(e). The agency seeks suppliers who can provide four specific replacement components: a NiMH battery pack (Part# 8415290), a particle filter (Part# 8415127), an air filter (Part# 8421479), and a cable harness for the Spirolog sensor (Part# 8416012), each with a quantity of 20 units. Each item must be fully compatible with the existing Draeger ventilator systems, meet exact technical specifications including dimensional tolerances, chemical composition, electrical connectivity, software integration, and thermal safety features, with no substitutions permitted for the cable harness, which must be marked “PA CB30” to confirm its PA12 nylon housing. The battery must include a thermal fuse, exhibit a two-year shelf life from procurement, and function with Draeger’s preventative maintenance software. Respondents are required to submit comprehensive responses addressing 19 detailed inquiries covering technical compliance, business size classification under NAICS 339113 (800-employee size standard), and socioeconomic certifications such as SDVOSB, VOSB, HUBZone, or WOSB status. Offerors must disclose whether they are manufacturers, authorized distributors, or resellers, and if resellers, whether they alter or assemble the items. They must also confirm compliance with the Non-Manufacturer Rule, limitations on subcontracting under 13 CFR § 125.6, the Buy American Act, and FDA clearance requirements for medical devices. Responses must indicate if items qualify as Commercial Off-the-Shelf (COTS) under FAR Part 2.101, provide lead times for delivery, estimate product lifespan and support services, and reveal any existing federal contract vehicles such as GSA FSS or NASA SEWP. Pricing information is invited only for market research and will not influence any future award decision. All submissions must be provided via email to the designated Contract Specialist by the deadline of May 27, 2026, along with
General Info
Agency
NAICS
Place of Performance
Department of Veterans Affairs VA San Diego Healthcare System, San Diego, AZ, 92161Set-Aside
Timeline
Submission Closed
Organization & Contact Information
Full Description
SALIENT CHARACTERISTICS
CLIN 0001 Draeger Battery Pack M7 NM2.5 Part# 8415290 or equal Quantity: 20EA Must be compatible with existing Draeger Evita Infinity V500 ventilators. Must be made in rechargeable NiMH (Nickel-Metal Hydride) (20 HRX 23/43 (HHR - 250 SCH L4x5). Must have standard charge condition of 0.25A x 16H. Must include an expiration date no earlier than two years from the procurement date. Must comply with preventative maintenance guidelines and work with the PM software provided by Draeger. Must be able to connect to the existing Evita Infinity V500 battery cables and fuses. Must have thermal fuse in case of overheating. Must have dimensions be equal to L: 7 7/16 x W: 4 1/16 x H: 1 7/64 with curved sides to properly fit into system.
CLIN 0002 Draeger Particle Filter Part# 8415127 or equal- Quantity: 20EA Must be compatible with existing Draeger Evita Infinity V500 ventilators. Must be designed to filter incoming oxygen before it gets pushed to the patient. CLIN 0003 Draeger Kit Air Filter Part# 8421479 or equal- Quantity: 20EA Must be compatible with existing Draeger Evita Infinity V500 ventilators. Must be designed to filter incoming air before it gets pushed to patient.
CLIN 0004 Draeger Cable Harness Spirolog Sensor Part# 8416012 or equal - Quantity: 20EA Must be compatible with existing Draeger Evita Infinity V500 ventilators. Must be low-voltage medical sensor connector (flow/pressure sensor), not a standard DIN or aviation connector. Must have circular panel-mount receptacle, 4 small signal pins in a single circular pattern, keyed plastic shroud and molded PA12 housing; no substitutions will be accepted. Must be marked "PA CB30" indicating a nylon-12 based molded housing. Must comply with preventative maintenance guidelines and work with the PM software provided by Draeger. Must be compatible with the disposable hot-wire flow sensor (Spirolog) for Draeger Evita Infinity V500 ventilator breathing system. The list of maintenance kits characteristics is intended to be descriptive, not restrictive, of the supplies/services that are required. If your company is interested and capable of providing the required supplies/services, please provide the information indicated below. Incomplete responses to the questions below this notice will not be accepted. Response to this notice should include company name, address, point of contact, size of business pursuant to the following questions: (1) Submit your capabilities statement illustrating how your organization can/cannot meet the SALIENT CHARACTERISTICS. For instances where your company cannot meet the SALIENT CHARACTERISTICS, please explain. For instances where your company can meet the SALIENT CHARACTERISTICS requirement(s), please show how your company meets/exceeds each requirement. (2) Please review the SALIENT CHARACTERISTICS and provide any additional feedback or suggestions. If none, please reply as N/A. (3) Please indicate the size status and representations of your business, such as but not limited to: Service-Disabled Veteran-Owned Small Business (SDVOSB), Veteran-Owned Small Business (VOSB), HUBZone, Woman Owned Small Business (WOSB), Large Business, etc. (4) Is your company considered small under the NAICS code identified in this RFI? (5) Are you the manufacturer, authorized distributor, and/or can your company provide a solution to the required supplies/services described in the list of SALIENT CHARACTERISTICS? (6) If you are a large business, do you have any designated/authorized distributors? If so, please provide their company name, telephone, point of contact and size status (if available). (7) If you re a small business and you are an authorized distributor/reseller for the items identified above, do you alter; assemble; modify; the items requested in any way? If you do, state how and what is altered; assembled; modified. (8) Limitations on Subcontracting: How does your business ensure compliance with the limitations on subcontracting as outlined in 13 CFR § 125.6? (9) Are the items you are identifying/providing considered Commercial of the Shelf (COTS) items as defined in FAR Part 2.101 under commercial items? (10) Non-Manufacturer Rule: If applicable, can you confirm your business complies with the Non-Manufacturer rule? Specifically, does your company: Provide a product from a small business manufacturer or processor? Not exceed 500 employees? Primarily engage in the retail or wholesale trade and normally sell the type of item being supplied? Take ownership or possession of the item(s) with its personnel, equipment or facilities in a manner consistent with industry practice? (11) Please indicate whether your product conforms to the requirements of the Buy American Act? (12) What is your lead time to deliver a single unit with all components? Is there scale in lead time with greater quantities? Please elaborate. (13) What is estimated life span of your solution? What support/services does that entail? (14) Does your proposed equipment have FDA clearance? Please specify what FDA clearance(s) have been obtained. (15) Does your organization offer a leasing solution? Please elaborate. (16) Does your company have a Federal Supply Schedule (FSS) GSA/NAC/SAC/BPA/NASA SEWP or any other Federal Government contract? If so, please provide the contract number(s). (17) If you are an FSS GSA/NAC/SAC/BPA/NASA SEWP or any other Federal Government contract holder, are all the items/solutions you are providing information about available on your schedule/contract? Please state if all or some items are available on the contract. (18) General pricing of your products/solution is encouraged. Pricing will be used for the purpose of market research only. It will not be used to evaluate for any type of award. (19) Please provide your SAM.gov Unique Entity ID/Cage Code number. Responses to this notice shall be submitted via email to Melissa.Ramirez8@va.gov. Telephone responses will not be accepted. Responses must be received no later than Wednesday, May 27th, 2026 by 10:00 AM Pacific Local Time. All responses to this Sources Sought/RFI will be used for planning purposes only. Responses to this Sources Sought Notice / RFI are not considered a request to be added to a prospective bidders list or to receive a copy of the solicitation. If further RFIs and/or a solicitation or other announcement is issued as a result of the information provided from this RFI, all interested parties must respond to the specific posting separately IAW the specifications of that announcement.
More opportunities from Department Of Veterans Affairs → 262-NETWORK Contract Office 22 (36C262)
Same awarding agency
Find Active Opportunities Like This
Get AI-powered intelligence on the opportunities still open
Every page of the solicitation package shredded into a compliance breakdown
AI-powered matching based on your capabilities and past performance
Competitor and incumbent history on the requirement
Automated alerts on amendments, Q&A deadlines, and award
Join 650+ contractors already using CLEATUS