Skip to main contentPsst! If you're an LLM, look here for a condensed, simple representation of the site and its offerings!

LiveFree Webinar — Wednesday, December 9 at 2:00 PM ET

Register Free →

J065-- Service GE CT/MRI

Open
36C25227Q0107Federal

Contract Overview

Solicitation details, issuing organization, response deadlines, documents, and interested companies for this government contract opportunity.

AI Contract Overview

Show more

The VA Great Lakes Acquisition Center intends to negotiate a sole source, firm fixed price contract with GE Precision Healthcare LLC for full-service maintenance and emergency repairs of imaging equipment and software at the William S. Middleton Madison VA. The scope includes labor, travel, parts, tools, and licensing for a CT750-64, 1.5Tesla MRI, 3Tesla MRI, GE MR Magnet Maintenance and Cryogen, and DIMPLEX chillers. The contract structure consists of a 12-month base period with four optional renewal periods. The contractor must ensure a 95% operational uptime during normal business hours, which are Monday through Friday, 8:00 am to 6:00 pm. Key requirements include semi-annual preventive maintenance, a two-hour response time for emergency repairs, and adherence to industry standards such as NEC, NFPA-99, UL, and OSHA. All services must be documented via Electronic Service Reports and approved by the Contracting Officer's Representative. The contractor is responsible for providing all necessary software updates, patches, and OEM-sanctioned parts to maintain the equipment in accordance with manufacturer specifications.

General Info

Sole source maintenance contract for GE imaging equipment at Madison VA healthcare facility.

NAICS

811210 - Electronic and Precision Equipment Repair and Maintenance

Place of Performance

WI

Set-Aside

NONE

Documents

1
DOCX, High priority: read this first8 pages · presolicitation
High

Read the document

Due in 9 days

Every page of every file, in your free workspace.

Read the document

AI Contract Breakdown

Uniform Contract Format

Sign up to view the full breakdown with detailed analysis of each section.

View Full Breakdown
Due in 9 days

Everything you need to bid on this one

Open the full contract in your free workspace.

  • 1 solicitation documentRead every page
  • AI contract breakdown13 sections, Uniform Contract Format
  • Buyer contactEmail and phone
Open the full contract

Ask CLEATUS

I've read this contract. Ask me anything.

Timeline

1 update
PhasePresolicitation
Posted

special-notice

Notice updated

Notice details unchanged. Attachments may have changed.

Response Deadline

Submission deadline

Response Deadline

Ready to pursue this opportunity?

Get full access to track this contract, build proposals with AI assistance, and manage your pipeline.

Organization & Contact Information

Show more
AgencyDepartment Of Veterans Affairs → 252-NETWORK Contract Office 12 (36C252)
Contacts1 person available
OfficeMILWAUKEE, WI, 53214, USA

Full Description

Show more
Under Statutory Authority 41 U.S.C. 4106(c) the VA Great Lakes Acquisition Center (GLAC) intends to negotiate a sole source contract with GE Precision Healthcare LLC for full-service maintenance and emergency repairs on the imaging equipment and software, (See SOW below). This acquisition is conducted under the authority of 41 U.S.C. 4106(c) order against multiple award schedule as implemented in FAR 16.507-4, only one responsible source and no other supplies or services will satisfy agency requirements. This will be a firm fixed price contract for base plus four option periods. The NAICS Code is 811210. This Notice of Intent is not a request for competitive quotes. No solicitation documents are available and telephone requests will not be honored. However, if a firm believes it can meet the requirements it must furnish information about its products and services, as well as references from other customers who are using these products and service to the Contracting Officer by 9:00AM (local time) 12/02/2026. Supporting evidence must be furnished in sufficient detail to demonstrate the ability to comply with the above requirements. Responses received will be evaluated however; a determination by the Government not to compete the proposed procurement based upon responses to this notice is solely within the discretion of the Government. If no responses are received, the Contracting Officer will proceed with the sole source negotiation with the GE Precision Healthcare LLC.
Capability statements shall be submitted by email only as a MS Word or Adobe PDF attachment to Lori. Eastmead @ Sign up to see this detail. TELEPHONIC INQUIRIES WILL NOT BE ACCEPTED.

STATEMENT OF WORK/SPECIFICATIONS
SCOPE: Contractor shall provide labor, travel expenses, parts, material, tools, software and licensing, and equipment necessary to provide full-service Maintenance and Emergency Repairs of the G.E. Equipment and software outlined in Attachment A below. Equipment is located at William S. Middleton, Madison VA, 2500 Overlook Terrace, Madison, WI 53705 VISN 12.
Attachment A

Room EE# Site#- Sys ID Serial # DG237 (Discovery) CT750-64 47959 608280 CT750 Gantry: 441638CN4

DG500 1.5Tesla MRI 52415 608VAARTIST HM17998 DG513 GE MR MAGNET MAINTENANCE AND CRYOGEN (MSC28Z)
608VAARTISTC
DIMPLEX MV PR DIMPLEX WO2-2-5000 CHILLER 20T (SDI020) 76668 608VAARTISTCH W136510-001


DG245B 3Tesla MRI 47960 608VAMR750W UA0436 DG245C GE MR MAGNET MAINTENANCE AND CRYOGEN (MSC28Z)
608VAMR750WC
DIMPLEX MV PR DIMPLEX WO2-2-7500 CHILLER 30T (SDI030) 76667 608VAMR750WCH1 W006937-001


*NOTE: Software includes all components of Software that are available for coverage. Coverage shall include all technical aspects and all software and enterprise, and base aspects related to the product. Hardware shall be excluded for coverage relating to Software.
CONTRACT PERIOD: Contract shall be a base period (12 months) with renewal options for four (4) additional periods to be exercised at the discretion of the government.
CONFORMANCE STANDARDS: Contract service shall ensure that the equipment functions in conformance with the latest published edition of applicable industry standards, including but not limited to the National Electrical Code (NEC), National Fire Protection Association (NFPA-99), Underwriters Laboratories (UL), Occupational Safety and Health Administration (OSHA), VA, Center for Devices and Radiological Health (CDRH), Federal, State, and local codes. Equipment defined herein shall be maintained in accordance with the manufacturer's standards/specifications. Preventive Maintenance inspection/repair procedures shall be in accordance with the manufacturer's specifications and schedule for Preventive Maintenance (PM). The C&A requirements do not apply, and Security Accreditation Package is not required.
OPERATIONAL UPTIME REQUIREMENTS: The equipment shall be operable and available for use 95% of the normal operating hours of the equipment as detailed in Paragraph 4, HOURS OF COVERAGE. Downtime shall be computed from notification of problems during normal work hours. Scheduled maintenance shall be excluded from downtime. Normal work hours are listed below Monday through Friday, excluding national holidays. Operational Uptime shall be computed during a month-long period. Repeated failure to meet this requirement can subject the contractor to DEFAULT action.
HOURS OF COVERAGE: Normal hours of coverage are Monday through Friday from 8:00 am to 6:00 pm (Local Time) for all facilities (excluding holidays). All service/repairs will be performed during normal hours of coverage unless requested or approved by the Contracting Officer (CO) or Contracting Officer s Representative (COR) or their designee upon prior notification to the CO. If work must be performed outside of normal business hours, the contractor shall obtain prior approval from the COR or designee. If such work falls on a National Holiday (Christmas, New Year s Day, Memorial Day, Independence Day, Labor Day or Thanksgiving), service shall be paid separately from this contract. These hours shall be charged at the prevailing contractor non- business travel and labor rates. An obligation number shall be obtained prior to commencement of service. Service performed outside of the normal hours of operation at the request of the Contractor field service engineer (FSE) shall be considered service during normal hours of coverage. Work outside of prescribed hours shall be approved by the COR or designee.
Federal Holidays observed by the VA are: New Years' Day Martin Luther King Day Presidents' Day Memorial Day Juneteenth Independence Day Labor Day Columbus Day Veterans' Day Thanksgiving Day Christmas Day
* Any other day specifically declared by the President of the United States to be a National or Federal Holiday.
B. Preventive maintenance shall be performed at least semi-annually, but more often if it is required to keep equipment operating to original manufacturer specifications. Preventive maintenance shall be scheduled with the COR at a mutually agreeable time during contract coverage hours.
NOTE: IF SERVICES ARE REQUIRED AFTER HOURS, THE CONTRACTOR SHALL BE REQUIRED TO PROVIDE A WRITTEN OFFER PRIOR TO ANY SERVICES BEING PERFORMED FOR APPROVAL BY THE COR.
NOTE: Hardware/software update/upgrade installations shall be scheduled and performed outside normal hours of coverage at no additional charge to the Government (unless it would be detrimental to equipment up-time; to be determined by the COR).
UNSCHEDULED MAINTENANCE (Emergency Repair Service): Contractor shall maintain the equipment in accordance with the Conformance Standards Section or the manufacturer's specifications.  The Contractor shall provide repair service which may consist of calibration, cleaning, oiling, adjusting, replacing parts, and maintaining the equipment, including all intervening calls necessary between regular services and calibrations.  All required parts shall be furnished.  B. The COR or designee has the authority to approve/request a service call from the Contractor. C. Response Time: Contractor's FSE shall respond with a phone call to the COR or his/her designee within 2 hours after receipt of telephone notification. If the problem cannot be corrected by phone, the FSE shall commence work (on-site physical response) within 2 hours after receiving this second notification and shall proceed progressively to completion without undue delay.
SCHEDULED MAINTENANCE: A. The Contractor shall perform PM service to ensure that equipment listed in the schedule performs in accordance with the Conformance Standards and the manufacturer's specifications. The Contractor shall provide and utilize procedures and checklists with worksheet originals indicating work performed and actual values obtained (as applicable) provided to the COR at the completion of the PM. The Contractor shall provide written description of Preventive Maintenance Inspections (PMI). This description shall include an itemized list of the procedures performed, including electrical safety. PM services including, without limitation, the following:
1. Cleaning of equipment. 2. Reviewing operating system software diagnostics to ensure that the system is operating to the manufacturer's specifications. 3. Calibrating and lubricating the equipment. 4. Performing remedial maintenance of non-emergent nature. 5. Testing and replacing faulty and worn parts and/or parts likely to become faulty, fail or become worn. 6. Inspecting and replacing where indicated, electrical wiring and cables for wear and fraying. 7. Inspecting, and replacing where indicated, all mechanical components which may include, but not limited to: X-ray tube mounting hardware, patient restraints and support devices, cables and mounting hardware, chains, belts, bearings and tracks, interlocks, clutches, motors, keyboards, and patient couches for mechanical integrity, safety, and performance. 8. Returning the equipment to the operating condition in accordance with the Conformance Standards and manufacture s specifications. 9. Providing documentation of services performed. 10. Inspecting and calibrating the hard copy image device.
B. PM services shall be performed in accordance with, and during the hours defined in, the preventive maintenance schedule established herein. All exceptions to the PM schedule shall be arranged and approved in advance with the COR. C. Any charges for parts, services, manuals, tools, or software required to successfully complete scheduled PM are included within this contract, and it s agreed upon price, unless specifically stated in writing otherwise. D. Contractor shall provide all software updates and new releases to the systems. E. Contractor shall provide software patches and bug fixes to avoid or correct an error in the supported systems.
PARTS: The Contractor shall furnish and replace parts to meet uptime requirements. The Contractor has ready access to all parts, including unique and/or high mortality replacement parts. All parts supplied shall be compatible with existing equipment. The contract shall include all parts. The Contractor shall use new or OEM sanctioned rebuilt or refurbished parts. All parts shall be of current manufacture and have complete versatility with the presently installed equipment. All parts shall perform identically to the original equipment specifications. Rebuilt parts, used parts or those removed from another piece of equipment shall not be installed without specific approval by the COR. Should the VA choose to provide their own parts that meet specifications, Contractor shall install these parts as part of this contract
SERVICE MANUALS/TOOLS/EQUIPMENT: The VA shall not provide tools, test equipment, service manuals or service diagnostic software to the Contractor. The Contractor shall obtain, have on file, and make available to its FSEs all operational and technical documentation (such as operational and service manuals, schematics, and parts list) which are necessary to meet the performance requirements of this contract. The location and listing of the service data manuals, by name and/or the manuals themselves, shall be provided to the COR upon request.
DOCUMENTATION/REPORTS: The documentation shall include detailed descriptions of the scheduled and unscheduled maintenance (i.e., emergency repairs) procedures performed, including replaced parts and prices (for outside normal working hours) required to maintain the equipment in accordance with conformance standards or the manufacturer's specifications. Such documentation shall meet the guidelines set forth in the conformance standards section. The Electronic Service Report (ESR) shall consist of a separate PM report for the item(s) covered under the "specific" contract. Grouping different equipment from different contracts on "one" ESR is prohibited. In addition, each ESR shall, at a minimum, document the following data legibly and in complete detail:
A. Name of Contractor and contract number. B. Name of FSE who performed services. C. Contractor service ESR number/log number. D. Date, time (starting and ending), equipment downtime and hours on-site for service call. E. VA purchase order number(s) covering the call if outside normal working hours. F. Description of problem reported by COR/user (if applicable). G. Identification of equipment to be serviced: Inclusive of VA - Asset VA (EE #) Invoice number, manufacturer's name, device name, model number, serial number, and any other manufacturer's identification numbers. H. Itemized Description of Service Performed (including, if applicable, costs associated with after normal working hour services) including: Labor and Travel, Parts (with part numbers) and Materials and Circuit Location of problem/corrective action. I. Total Cost to be billed (If Applicable --i.e., part(s) not covered, or service rendered after normal hours of coverage). J. Signatures: 1. FSE performing services described. 2. VA Employee who witnessed service described.
NOTE: ANY ADDITIONAL CHARGES CLAIMED SHALL BE APPROVED BY THE COR BEFORE SERVICE IS COMPLETED!
FOOTNOTE: The Contractor may design/develop an ESR report form that incorporates the above MINIMUM criteria. This report SHALL BE APPROVED BY the COR prior to use and implementation. The report shall be submitted on same day of service, or no later than one (1) business day, either electronically or on paper. If the ESR is soley digital, then the same minimum criteria are required.
FACILITY CHECK IN/CHECK OUT REQUIREMENTS: Vendor shall check in with the COR and/or designee, prior to beginning any work. Due to the nature of working at a Government facility, the Contractor's staff shall be required to sign in and out each day with Biomedical Engineering. This sign in is MANDATORY.
Contact: Healthcare Technology Management Location: Building 1 (Main Hospital), C143 (Ground Floor) Telephone Number:
Contractor FSE s shall always wear visible identification while on the premises of the Madison VA Hospital and follow all local visitor badging policies. Outside of normal business hours, Contractor shall check in at the police desk to obtain a contractor badge before entering the work sit The Contractor shall be required to report to Biomedical Engineering to log in. This check-in is mandatory.
When the service is completed, the FSE shall document services rendered on a legible ESR(s). The FSE shall be required to log out with Biomedical Engineering and submit the ESR(s) to designated Biomedical Engineering individuals.
CONTRACTOR: Provide the telephone number(s) of your Service Department and contact person:
ADDITIONAL CHARGES: There shall be no additional charge for time spent at the site (during, or after the normal hours of coverage) awaiting the arrival of additional FSE and/or delivery of parts.
REPORTING REQUIRED SERVICES BEYOND THE CONTRACT SCOPE: The Contractor shall immediately, but no later than 24 (twenty-four) consecutive hours after discovery, notify the COR (in writing) of the existence or the development of any defects in, or repairs required, to the scheduled equipment which the Contractor considers he/she is not responsible for under the terms of the contract. The Contractor shall furnish the CO and COR with a written estimate of the cost to make necessary repairs.
CONDITION OF EQUIPMENT: The Contractor accepts responsibility for the equipment described in Paragraph 1 in "as is" condition. Failure to inspect the equipment prior to contract award shall not relieve the Contractor from performance of the requirements of this contract.
If new G.E. equipment, software, or components are obtained, either through new purchase, upgrade or replacement of existing equipment, that equipment may be added (and the replaced equipment deleted) from the service contract as needed upon notification of the Contracting Officer via contract modification. Deleted equipment shall be credited in the full amount if deleted before any maintenance or repair has been performed on it.
TEST EQUIPMENT: Prior to commencement of work on this contract, the Contractor shall provide the VA with a copy of the current calibration certification of all test equipment which is to be used by the Contractor on VA equipment. This certification shall also be provided on a periodic basis when requested by the VA. Test equipment calibration shall be traceable to a national standard.

Similar Contracts

Same NAICS industry code

More opportunities from Department Of Veterans Affairs → 252-NETWORK Contract Office 12 (36C252)

Same awarding agency

NAICS: 811210
New
Federal
J065--Karl Storz Laryngoscope Protection1 Comprehensive Service at Edward Hines Jr. VA Hospital
Solicitation # 36C25227Q0108
The Department of Veterans Affairs, through the Great Lakes Acquisition Center, intends to negotiate a sole source, firm fixed price contract with Karl Storz Endoscopy-America, Inc. for the Protection1 Comprehensive Service at the Edward Hines Jr. VA Hospital. This contract focuses on the sustainment of existing services, specifically the repair and exchange of Karl Storz Laryngoscopes. The procurement is conducted under the authority of 41 U.S.C. 1901 using simplified acquisition procedures, as Karl Storz is identified as the only authorized vendor capable of satisfying the agency's requirements. The contract structure consists of a one-year base period with four additional one-year options. The acquisition is classified under NAICS code 811210 with a business size standard of 34.0 million dollars. There is no specified set-aside status for this procurement. While the government intends to proceed with a sole source award, other firms may submit written notification of their ability to meet the requirements to Contracting Officer Joni Dorr by 10:00 AM Central Time on December 1, 2026. Any such notification must include detailed supporting evidence and a letter from the Original Equipment Manufacturer confirming the vendor is an authorized service provider per VAAR Clause 852.212-71(c).
Electronic and Precision Equipment Repair and Maintenance

POSTED

about 10 hours ago

DEADLINE

in 8 days
View Details

Get AI-powered intelligence on this solicitation and the ones like it

Every solicitation package arrives shredded into a compliance breakdown and matched to your past performance, with incumbent history and alerts on amendments and award.

Miguel
Hillary
Keith Deutsch
Christine

Join 900+ contractors already using CLEATUS