Guldmann Patient Ceiling Lift - Marion IL

Closed
36C25526Q0472Federal

Contract Overview

Solicitation details, issuing organization, response deadlines, documents, and interested companies for this government contract opportunity.

Active Opportunities Like This One

NAICS: 339112
New
Federal
ScriptPro - Pharmacy Workstations: Kiosk and Med Dispensing Unit
Solicitation # 36C24W27Q0012
The Department of Veterans Affairs Veterans Health Administration is conducting market research via a Sources Sought notice for the procurement of brand name ScriptPro pharmacy workstations, kiosks, and medication dispensing units for the Oakland VA Health Care System in California. This is a Firm Fixed Price requirement intended to identify capable manufacturers or distributors who can provide exact match or functionally equivalent equipment. The scope includes the purchase, delivery, assembly, and installation of three pharmacy workstations, three Zebra ZD621-2P printers, three Rx paperwork printers, one pharmacy patient kiosk, one Eyecon Model 9430, and specific pharmacy server software. To ensure compatibility with existing systems, the requirement is designated as brand name only, though equal products may be considered if they meet all salient physical and functional characteristics. The government strictly prohibits gray market items, requiring vendors to be OEMs or authorized dealers/distributors with valid authorization letters. Contractors must adhere to rigorous security standards, including FIPS 140-2 and NIST 800-52 compliance, and utilize VA-authorized technical security baseline configurations. Deliveries must be coordinated with 24-hour advanced notice and may involve working within active construction sites, requiring compliance with safety protocols such as the use of hard hats and safety vests. Performance will occur during normal business hours at the Oakland VA Clinic, and the contractor is responsible for all tools, labor, and materials necessary for a complete installation. This notice is for planning and size determination purposes only and does not constitute a formal solicitation or binding contract.
Rpo West (36C24W)

POSTED

about 5 hours ago

DEADLINE

in 6 days
NAICS: 339112
New
ELECTRODE — 5977004028529 — N0010426RFG70
Solicitation # N0010426RFG70
Solicitation N00104-26-R-FG70 is a negotiated procurement issued by NAVSUP Weapon Systems Support for 500 units of a Plug-In Type Electrode (NSN 5977-00-402-8529). This material is designated as Special Emphasis Level I due to its use in crucial shipboard systems where failure could result in serious personnel injury or loss of the vessel. The contract requires a destructive First Article Test (FAT) to evaluate the manufacturer's capability prior to full production. Award will be made based on the lowest priced, technically acceptable offer, with a final delivery timeline of 186 days after the effective date of the contract. The technical requirements are stringent, mandating compliance with Naval Ship Systems Command Drawing 815-1197212 and MS16662, alongside specific material certifications for chemical and mechanical analysis of the body, upper electrode, and nut. Quality assurance is managed through Government Source Inspection, requiring a quality system compliant with ISO-9001 or MIL-I-45208. Critical requirements include 100% inspection of material verification and traceability data, strict adherence to FED-STD-H28 for thread inspection, and the submission of certification data via Wide Area WorkFlow (WAWF) at least 20 days prior to delivery. All items must be mercury-free and packaged according to MIL-STD-2073 and MIL-STD-129.
Navsup Weapon Systems Support

POSTED

about 13 hours ago

DEADLINE

in about 1 month
NAICS: 339112
New
DIBBS
MASK, ORONASAL
Solicitation # SPE2DS-27-T-1019
Solicitation SPE2DS-27-T-1019 is a request for quotations from the Department of Defense, specifically the Medical Supply Chain MD Surg FSF, for the procurement of universal size oronasal masks under NSN 6515-01-276-1417. The required items feature a head strap coupling and a pre-inflate cushion to prevent mouth-to-mouth contact during resuscitation and must be provided in a pocket-size case with illustrated instructions. The procurement consists of two line items, each for a quantity of one unit, with delivery required within 20 days after receipt of order. Delivery is set as FOB Destination, with shipping addresses including the USNS Burlington T-EPF 10 and NAVSUP FLC North New London. The contract mandates strict adherence to quality and packaging standards, including ASTM D3951, MIL-STD-129 for marking and labeling, and RP001 for palletization, with the DLA Master List of Technical and Quality Requirements taking precedence. Inspection and acceptance will occur at the destination per FAR 52.246-2. Administrative requirements include the use of the Wide Area Workflow system for electronic invoicing and payment. Additionally, the award is subject to various federal and defense regulations, including the Buy American Act, the Berry Amendment, and specific cybersecurity reporting requirements under DFARS 252.204-7012. All quotes must be submitted through the DLA Internet Bid Board System.
MEDICAL SUPPLY CHAIN MD SURG FSF

POSTED

about 24 hours ago

DEADLINE

in 5 days

AI Contract Overview

Show more

The Marion VA Health Care System is seeking a contractor to provide comprehensive turn-key installation of permanent ceiling-mounted patient lift systems across existing facilities in Marion, Illinois, with a focus on modernizing outdated infrastructure built between the early 1900s and 1950s that feature limited ceiling clearance and complex utility layouts. The scope includes full design, site evaluation, removal of existing systems, installation of new ceiling lift systems, and a buy-back proposal for decommissioned equipment. All new lifts must have a minimum lifting capacity of 770 lbs, operate with smooth, controlled motion, and be designed as X-Y room-covering rail systems to maximize space utilization. Installation must not rely on existing hangers, Unistrut, or utility supports; all structural attachments must connect solely to building load-bearing components. The contractor must submit detailed room-specific plans, mechanical anchor designs, anchoring drawings to structural concrete, and a phasing installation schedule that minimizes disruption to patient care, prioritizing low-traffic periods such as weekends or off-peak days. All work must comply with VA Seismic Guidelines (H18-8), International Building Code IBC 2006, National Electric Code, Life Safety Code, Joint Commission standards, and ISO EN 60601-1 and ISO EN 10535 for medical and patient handling equipment. Door frames may not be modified unless a licensed structural engineer certifies the changes and the VA approves them. Technical requirements mandate that all new lift systems undergo 150% weight testing at all suspension points and traverse rails, with deflection measurements documented and verified by the VA COTR before acceptance. The contractor must provide written certification of seismic compliance for each installation and complete standardized checklists from VACO Patient Safety Alert AL14-07PSA for installation, relocation, and preventive maintenance, signed by the technician, facility representative, and unit manager. All personnel performing maintenance or installation must be factory-certified by Guldmann and adhere to Universal Precautions when handling equipment with potential bloodborne pathogen exposure, using personally provided PPE. Dust containment and post-installation cleaning are required, including tacky mats at entrances and thorough vacuuming or wet wiping of all surfaces. A two-year warranty on equipment and five-year warranty on installation must be provided, alongside an extensive training program for all staff including biomedical personnel on usage, inspection, and annual weight testing procedures. The contractor must submit a detailed follow-up support plan covering repairs,

General Info

Install ceiling patient lift systems with Guldmann certification, seismic compliance, 770 lb capacity, and full VA-regulated turn-key services.

NAICS

339112 - Surgical and Medical Instrument Manufacturing

Place of Performance

Marion, IL, 62959, USA

Set-Aside

NONE

Documents

1

Sources Sought Notice for Guldmann Patient Ceiling Lifts Installation - Solicitation 36C25526Q0472

PDF, High priority: read this firstsources-sought
High

AI Contract Breakdown

Uniform Contract Format

Sign up to view the full breakdown with detailed analysis of each section.

Timeline

PhaseClosed
Posted

Sources Sought

Response Deadline

Deadline has passed

Submission Closed

Find active opportunities like this

Get full access to discover similar active contracts, track opportunities, and build proposals with AI assistance.

Organization & Contact Information

Show more
AgencyDepartment Of Veterans Affairs → 255-NETWORK Contract Office 15 (36C255)
Contacts1 person available
OfficeLEAVENWORTH, KS, 66048, USA
Office AddressLEAVENWORTH, KS, 66048, USA

Full Description

Show more

Marion, IL


Statement of Work


Patient Ceiling Lifts



Background:


The Marion VA Health Care System is seeking a contractor to furnish all design, site evaluation, labor, materials, and equipment to install permanent ceiling mounted patient care lifts in existing facilities. Many hospital buildings were constructed in the early 1900-1950’s and have limited above ceiling clearance. Additionally, numerous utility systems occupy this space and must be considered when installing new equipment above the ceiling. It is strongly recommended that the contractor perform a thorough site evaluation prior to submitting a proposal.



Scope of Work:


The Contractor shall provide a technical and cost proposal for turn-key services to install patient lift systems in accordance with this scope of work. This cost proposal shall be for providing installation of all new and removal/de-installation of all lift system equipment including but not limited to; slings, lift devices, rails, support material, bracing and all other associated components required for complete, functional installation for ceiling lift in with scale and continuous charge ceiling lift systems. This proposal is also a request for the vendor to provide a buy-back proposal for any old systems that are removed. This proposal shall also be for installation of the lift systems including but not limited to any required plans, engineering details and documents, tools, labor, testing, training, cleaning supplies, containment systems, etc.  All new lift systems must have a lifting capacity of no less than 770 lbs.  The rail systems must be designed, constructed, and installed in an X-Y (room-covering) design; in such a manner as to allow for maximum utilization of patient room space. Operation of the system must be smooth with no fast or jerky starts and stops.  Operation of the system must allow for adequate lifting height allowing for transfer of patient from bed.  Track systems shall be self supporting either from the ceiling or from a post mounted system as required due to existing conditions. Existing hangers, Unistrut, or other support components shall not be utilized to support the lift tracks. No existing utility systems or supports shall be used for support or for bracing of the lift system. All new supports must be connected directly to building structural components.






Technical Requirements:



  1. Contractor shall provide a plan for each room detailing location of lift rails, support structure above, and means of attachment.
  2. The lift systems must meet or exceed seismic bracing requirements as outlined in the VA Seismic Guidelines (H18-8). A written report certifying compliance with this standard shall be provided to the facility for each lift installation.
  3. Door frames are not to be cut or modified. If contractor proposes to cut or modify an existing door frame, documentation supporting this practice and providing modification procedures must be provided and certified by a licensed structural engineer and accepted by the VA.
  4. All lifts shall be weight tested prior to acceptance. Test weight shall be 150% of rated capacity and shall be applied to all points of suspension and traverse rail.  Prior to testing, contractor shall submit rail deflection testing standards to the VA. Testing shall be observed by the VA COTR and a written report shall be provided following each test. The report shall include deflection measurements taken at locations of maximum deflection during the weight test. All deflection points shall be within defined code limitations.

Certified contractor will not perform any work/preventive maintenance inspections in an area/room occupied by a patient. Unless otherwise approved by contract COR (or designee), certified contractor will be required to perform all work during the hospital normal business hours (M-F 8:00 AM – 4:30 PM) excluding federal holidays. Contractor will provide all travel, labor, parts, test equipment (to include weight set), tools, and incidentals necessary to perform all required inspection services. Any overtime needed that would cause an increasing in invoicing must receive prior approval by the VISN Contracting Officer, through the COR (or designee).


All services for this contract must meet manufacturer's performance and technical specifications, Federal Regulations, and other specifications that may apply such as National Electric Code (NEC), Life Safety Code (LSC), Joint Commission (JC), and VA Regulations. There is a potential for exposure to blood borne or other infectious material with equipment throughout the hospital.


All contractor maintenance persons must use the “Universal Precautions” during cleaning and maintenance/repair procedures required by this service request. Contractor will be responsible for providing their own Personal Protective Equipment as applicable.


A Guldmann certified contractor must annotate all aspects of annual inspection performed on the Guldmann annual inspection checklist to include: date and time of completed preventive maintenance; type, model, and serial number of the lifting system on which preventive maintenance was performed, maintenance performed and all repairs and/or repair parts (if applicable) used to maintain operating efficiency per manufacture specifications of the equipment. The Guldmann contractor will also complete checklists for Installation or Relocation Checklist for Ceiling Mounted Patient Lifts and/or Corrective and Preventive Maintenance Checklist for Ceiling Mounted Patient Lifts as outlined in VACO Patient Safety Alert AL14-07PSA. The applicable checklist must be completed and signed by the inspector/technician, the facility representative, and the unit manager where the work is done. The annual inspection checklist along with the VA inspection checklists (as applicable) shall be sent to the applicable facility POC (BMET, Facilities, or SPHM Facility Coordinator).


  1. Final equipment certification shall be provided in writing prior to acceptance by the VA.
  2. Any lift installation that modifies the function of a room’s privacy curtains shall be addressed by lift installer. The mitigation plan must be submitted to and approved by the VA.
  3. Prior to a manufacturer mismatch between lift motor and rails both manufacturers must have submitted in writing to the Contracting Officer that the rail system and lift assemblies are compatible and supportable, including the power and charging mechanisms, and provide testing results; the lift assembly manufacturer agrees to do preventive maintenance according to the checklist contained within Patient Safety Alert AL14-07 (Appendix B) and their own procedures; and the lift assembly manufacturer accepts all liability and warranty related to the lift assembly and the rail system structure once they have installed their lift assemblies on the rails.
  4. Ceiling lift manufacturers must allow and warrant the use of their lifts with the full range of Guldmann and EMR Distributors loop slings.
  5. Lift installation shall include dust containment during installation. All rooms shall be returned to the VA in their original condition. Any damage inflicted shall be corrected by the contractor. Contractor shall provide tacky mats at each entrance to the area where work is being done. Upon completion of installation all surfaces shall be vacuumed and/or wet wiped to remove dust/debris.  Guldmann certified contractor will be required to perform a complete and comprehensive Preventive Maintenance Inspection on any/all patient overhead lifting systems in any VISN 15 facility that will meet and comply with original equipment manufacturer certification specifications and also include deflection testing as required in Veterans Affairs Central Office (VACO) Patient Safety Alert AL14-07PSA Checklists for Installation or Relocation Checklist for Ceiling Mounted Patient Lifts and/or Corrective and Preventive Maintenance Checklist for Ceiling Mounted Patient Lifts.


Contractor shall provide the following to be considered for award:



  1. Disclosure of all Tort claims against vendor either pending or resolved regarding proposed system.
  2. Education/Training Plan by vendor on use of all equipment, after installation completion, with specific time lines -- all shifts, all areas and possibly refresher courses.  Include training plan for Bio-Medical staff which details any annual weight testing/inspection requirements required system warranty/certification
  3. Standardization of location of charging units in identified rooms.
  4. Written plan for installation of track lift system as applicable including mechanical anchor designs.
  5. Written plan for modifying existing structural, mechanical, electrical and other infrastructure systems if necessary.
  6. Provide drawings to show how rail system will be anchored to the structural concrete.
  7. Provide a specific time line for installation. 
  8. Provide an installation Infection Control Plan that meets the facility’s infection control risk assessment requirements.
  9. Provide an Installation Plan including phasing and indicating time per unit with minimum interference with patient care.  Example: installed on days when patient population is down, possibly weekend or Monday and Friday.  The least amount of disruption to patient’s comfort and care is expected. 
  10. A two year warranty on equipment and five year warranty on installation shall be provided to each facility.
  11. Specify all previous work done by vendor in VISN 15.  Is it complete, including all training?
  12. Provide Plan of Equipment Certification after install. 
  13. Provide Plan reflecting availability of vendor for follow- up, to include but not limited to: repairs, adjustment, sling replacement after installation is complete; and address after hours and weekend response during warranty period.
  14. Initial set up must include expendable accessories, such as slings. Provide one non-disposable, or two disposable, slings per lift, as per facility specifications. 
  15. Shall submit enough product information to allow the Government to evaluate the product(s) being offered. 
  16. Equipment installed shall be subject to the following Specs:

  • Available  as UL listed full-system installations (UL 60601-1)
  • Up to 26’ span for unspliced rail
  • 23’4” traverse rail span capability at 770 lbs.
  • 20’ traverse rail span capability at 1,000 lbs.
  • Supplier system engineered lateral bracing brackets (Not after market)
  • Optional InfraRed remote control of lift available
  • Optional InfraRed remote control of lateral movement for motor and traverse rail
  • O.S.H.P.D. (California seismic resistance) approved system installations available
  • Rust/stain-resistant, single-bar-design hangar bar with integrated bearing-swivel and closed-cell

handgrip; available in multiple sizes


  • 100% motor testing before shipment
  • In-field-serviceable software-based controller boards
  • Anodized aluminum rail finish to prevent flaking, chipping, fading, or peeling
  • Over-speed safety governor on motor
  • Soft start/stop control
  • Visual low battery indicator
  • Battery charger is splash proof and double insulated, provides continuous charge at all times, in any location within overhead rail system
  • Sound level of lifting motor to be less than or equal to 52dB(A)
  • Emergency switch to offer disengagement then emergency lower at controlled speed
  • All installations are completed by Factory Certified, US installers.
  • Available  UL certification of entire installation; as opposed to UL listing on the transformer only
  • Provision of system approval on room covering according to ISO EN 60601-1 Medical Equipment
  • Provision of system approval on room covering according to ISO EN 10535 Safe Patient Handling Equipment
  • Installation hardware meets IBC 2006 (International Building Code) standards


Payment will be made monthly in arrears upon receipt of a properly prepared invoice, referencing all lifting systems that received the Required Services the contract number and assigned purchase order number.



Personnel Qualifications


Contractor personnel performing preventive maintenance inspections and repair services must be fully qualified as per the original manufacture technical certification requirements.


Listing of equipment to be covered under the Required Services as stated in paragraph1 above.



More opportunities from Department Of Veterans Affairs → 255-NETWORK Contract Office 15 (36C255)

Same awarding agency

NAICS: 811210
New
Federal
J061--Uninterruptible Power Supply St. Louis VA Medical Center
Solicitation # 36C25526Q0678
The Department of Veterans Affairs, Network Contracting Office 15, is soliciting quotes for a five-year Indefinite Delivery Indefinite Quantity (IDIQ) firm-fixed-price contract to provide preventive maintenance, emergency services, repairs, and battery replacements for Uninterruptible Power Supply (UPS) units and batteries. The services will be performed at the St. Louis VA Health Care System, specifically at the John Cochran and Jefferson Barracks VA Medical Centers. The contract has a guaranteed minimum of $43,000.00 and a total ceiling of $215,000.00, with a period of performance extending from late 2026 through 2031. While initially a small business set-aside, the solicitation was updated via Amendment 0003 to be unrestricted. Award will be based on the Lowest Price Technically Acceptable (LPTA) process, where quotes are rated as either technically acceptable or unacceptable based on their ability to meet the Performance Work Statement and specifications. Contractors must provide evidence of safety training in OSHA 1910 Subpart S and NFPA 70E, as well as manufacturer certifications for the equipment being serviced. All invoices must be submitted electronically through the Tungsten system. The final deadline for quote submission is October 29, 2026, at 7:00 AM CDT.
Electronic and Precision Equipment Repair and Maintenance

POSTED

about 5 hours ago

DEADLINE

in 2 days
View Details
NAICS: 541620
New
Federal
F108--657A4-21-800 - VPIH Services for Remove and Replace End of Life Heating, Ventilation and Air Conditioning Construction
Solicitation # 36C25527R0004
The Department of Veterans Affairs, Network Contracting Office 15, is seeking qualified Service-Disabled Veteran Owned Small Business (SDVOSB) Architecture and Engineering (A/E) firms to provide VA Professional Industrial Hygienist (VPIH) services. This project supports the removal and replacement of end-of-life HVAC systems in Building 1 at the John J. Pershing VA Medical Center in Poplar Bluff, Missouri. The scope of work involves providing all labor, materials, equipment, and supervision for the oversight of asbestos and lead abatement activities, including Phase Contrast Microscopy (PCM) air monitoring and analysis, compliance monitoring, inspections, and clearance testing. The project is divided into four phases ranging from 8 to 14 days each. Key deliverables include the submission of all routine and clearance air sample results to the Contracting Officer's Representative within 24 hours of collection and a final report for each location summarizing pre- and post-abatement activities. This procurement is conducted under the Brooks Act and requires the submission of SF 330 qualification packages. Evaluation criteria focus on the resumes of the Project Manager and Quality Control Manager, specialized experience in active healthcare facilities, the firm's capacity to accomplish the work, and past performance. The anticipated contract type is Firm Fixed Price, and the updated NAICS code for this effort is 541620 for Environmental Consulting. Awarded contractors must comply with E-Verify and limitations on subcontracting, ensuring no more than 50 percent of the government-paid amount is paid to non-SDVOSB firms.
Environmental Consulting Services

POSTED

1 day ago

DEADLINE

in 5 days
View Details
NAICS: 561730
New
Federal
S208--Grounds Maintenance Services for the St. Louis VA Medical Center Jefferson Barracks
Solicitation # 36C25527Q0023
Solicitation 36C25527Q0023 is a request for quotations for grounds maintenance services at the 126.65-acre Jefferson Barracks campus of the St. Louis VA Medical Center. This five-year IDIQ contract, with a guaranteed minimum of 2,500 dollars and a maximum ceiling of 950,000 dollars, requires the contractor to provide all labor, equipment, and materials for routine maintenance, including mowing, trimming, edging, weed control, and seasonal cleanup. The scope also includes spring and fall fertilizer and pest treatments, as well as emergency landscaping services with a required four-hour response time. Performance is measured against a Quality Assurance Surveillance Plan with strict standards for grass height and debris removal. The procurement utilizes a tiered set-aside evaluation process, prioritizing Service-Disabled Veteran-Owned Small Businesses (Tier 1), followed by Veteran-Owned Small Businesses (Tier 2), and other small businesses (Tier 3), with further priority given to HUBZone, 8(a), and women-owned concerns. Award will be based on the best benefit to the government, evaluating experience, technical approach, and price. Technical evaluations focus on equipment availability and the proposed operational plan. Offerors must submit three separate files containing pricing and certifications, a record of up to five relevant contracts from the last five years, and a detailed technical submission. Personnel are required to have completed the OSHA 10-Hour Construction Safety Course, and the contract mandates compliance with various FAR clauses and safety regulations.
Landscaping Services

POSTED

1 day ago

DEADLINE

in 9 days
View Details
NAICS: 236220
New
Federal
Z2DA--657-24-111JC Renovate Ambassador Suite & Main Lobby B1
Solicitation # 36C25526R0167
Solicitation 36C25526R0167 is a Firm Fixed Price construction project issued by the Department of Veterans Affairs, Network Contracting Office 15, for the renovation of the Ambassador Suite and Main Lobby in Building 1 at the John Cochran VA Medical Center in Saint Louis, Missouri. This project is a 100% set-aside for Service-Disabled Veteran-Owned Small Businesses (SDVOSB) under NAICS code 236220, with an estimated value between 1 million and 5 million dollars. The scope of work includes improving ADA access, repairing skylight leaks, updating finishes, correcting HVAC and electrical deficiencies, and replacing the revolving door with sliding or swinging doors. The project must be completed within a performance period of 330 calendar days, though some technical documents specify a construction period of 218 days. Award will be based on the Lowest Price Technically Acceptable (LPTA) selection process. To be considered technically acceptable, offerors must demonstrate relevant experience with hospital construction projects of similar magnitude completed within the last five years and provide a qualified Site Superintendent and Safety Officer, both with at least five years of experience. Proposals must include a detailed construction safety plan for working in an active medical facility and a certification of compliance regarding subcontracting limitations, ensuring no more than 85 percent of the government payment is paid to non-SDVOSB or non-VOSB firms. Proposals are due by November 4, 2026, and must include a technical proposal, a price proposal via SF 1442, and a bid guarantee of at least 20 percent of the bid price, not to exceed 3 million dollars.
Commercial and Institutional Building Construction

POSTED

2 days ago

DEADLINE

in 7 days
View Details
NAICS: 236220
New
Federal
657-26-602 Create Observation Unit JC
Solicitation # 36C25526R0162
Solicitation 36C25526R0162 is a negotiated request for proposal issued by the Department of Veterans Affairs Network Contracting Office 15 for the construction of an Observation Unit at the Saint Louis VA Medical Center, John Cochran Division. This project, identified as 657-26-602JC, involves the conversion of three rooms into a 3-bed observation unit and the renovation of seven additional triage and reception rooms within the existing Emergency Department. The estimated construction magnitude is between 500,000 and 1,000,000 dollars, with a required period of performance of 550 calendar days from the issuance of the Notice to Proceed. The project is set aside specifically for Service Disabled Veteran Owned Small Businesses (SDVOSB). The contract will be awarded using the Lowest Price Technically Acceptable (LPTA) process, evaluating five technical factors including relevant past performance in 24-hour medical facilities, contractor and key staff experience, subcontractor qualifications, and a comprehensive safety plan. Following multiple amendments to incorporate revised bid documents, full drawing sets, and final responses to requests for information, the proposal due date has been extended to November 6, 2026, at 2:00 PM CST. Requirements include the submission of a technical proposal, a price proposal, and original bid bonds. Awardees must provide performance and payment bonds within 10 days of award and adhere to a one-year guarantee period for elevator services.
Commercial and Institutional Building Construction

POSTED

2 days ago

DEADLINE

in 9 days
View Details

Find Active Opportunities Like This

Get AI-powered intelligence on the opportunities still open

Every page of the solicitation package shredded into a compliance breakdown

AI-powered matching based on your capabilities and past performance

Competitor and incumbent history on the requirement

Automated alerts on amendments, Q&A deadlines, and award

Miguel
Hillary
Keith Deutsch
Christine

Join 750+ contractors already using CLEATUS