J061--Medical Gas Alarm Cable Services- MRN - MC 657-26-3-2009-0308 (VA-26-00078957)

Closed
36C25526Q0611Federal

Contract Overview

Solicitation details, issuing organization, response deadlines, documents, and interested companies for this government contract opportunity.

Active Opportunities Like This One

NAICS: 238210
New
Transformer Oil Testing Services
Solicitation # RFP-VC-26-010
The Inland Empire Utilities Agency (IEUA) is soliciting proposals under RFP-VC-26-010 for professional transformer oil testing services to enhance the reliability, safety, and maintainability of its transformers. The scope of work is divided into three primary areas: conducting visual and infrared thermographic inspections on energized transformers under typical load, installing standardized external oil sampling devices to improve safety and access, and performing comprehensive oil testing analysis. Additionally, the contractor is responsible for evaluating nitrogen blanket systems to ensure positive pressure and providing on-site training for up to 20 agency staff members on inspection tools and defect identification. All work must adhere to strict industry standards, including ASNT Level 1 certification for thermography, ASTM D923 for oil sampling, and NFPA 70E safety protocols. The agency intends to make a single award to the most capable offeror based on a cost-effective and time-efficient proposal. Key evaluation criteria include a proven track record of at least two similar projects within the last five years, the proposed staffing plan, personnel availability, and the fee structure. Proposals must be submitted electronically via PlanetBids by December 23, 2026, and must include a detailed project approach, company profile, and required certifications such as a Non-Collusion Statement and Workers' Compensation Certificate. The resulting contract will utilize fixed pricing with a not-to-exceed total value and may include one-year extension options at the agency's discretion. Payment will be processed electronically via ACH within 30 days of invoice approval by the Project Manager.
Inland Empire Utilities Agency

POSTED

about 1 hour ago

DEADLINE

in 2 months
NAICS: 238210
New
Purified Water Replenishment San Jacinto Valley Advanced Water Purification Facility
Solicitation # purified-water-replenishment-san-jacinto-valley-advanced-water-purification-facility-2
Axia Electric LLC is soliciting bids from SBE/DBE certified firms for the Purified Water Replenishment San Jacinto Valley Advanced Water Purification Facility and Brine Ponds project in California for the Eastern Municipal Water District. The requested scopes of work include low voltage systems, fiber optics, and the supply of electrical materials. Interested subcontractors must submit their price quotes by 2:00 PM on November 18th, with a final bid deadline of November 19th, 2026. Bidders are encouraged to provide information regarding any lower-tier SBE/DBE participation to be considered during the evaluation process. Qualified bidders must be registered to perform public work pursuant to Labor Code Section 1725.5 and must adhere to the AGC Master Labor Agreements. Awardees will be bound by the prime contract terms and Axia's Standard Long Form Subcontract or Purchase Order. Minimum insurance requirements include 1 million dollars for Auto Liability and Workers Compensation, 5 million dollars for Excess Liability, and General Liability coverage of 1 million dollars per occurrence, 2 million dollars aggregate, and 2 million dollars for completed operations. Primary points of contact for the solicitation are Ryan Falk for estimating and Amy Roller for contract documents.
Axia Electric

POSTED

about 1 hour ago

DEADLINE

in 28 days
NAICS: 238210
New
Centennial JCC Emergency Lighting Replacement, Upgrade, Testing & Maintenance Services
Solicitation # centennial-jcc-emergency-lighting-replacement-upgrade-testing-maintenance-services
Management and Training Corporation is soliciting bids for a fixed-price, single lump sum contract to provide emergency lighting replacement, upgrade, testing, and maintenance services at the Centennial Job Corps Center in Nampa, Idaho. The project involves a qualified electrical contractor inspecting, testing, repairing, and certifying emergency lighting systems to ensure compliance with NFPA 101 (2024 Edition) and NFPA 70 standards. Key deliverables include a full inventory of fixtures, functional testing via simulated power outages, the application of durable tracking labels to each fixture, integration of maintenance requirements into the facility's Computerized Maintenance Management System, and the provision of disposal certificates for hazardous materials. The estimated contract value ranges between $25,000 and $100,000. Bids are due by December 1, 2026, and must be submitted via email with a detailed cost breakout and the official MTC Bid for Lump Sum Contract form. A mandatory physical site visit is required prior to bid submission. For bids equal to or greater than $25,000, contractors must provide a bid bond guarantee of 20% of the base bid, as well as performance and payment bonds from a surety company with at least an A- rating. The award will be based on overall best value. This opportunity is set aside for small businesses, including certified VOSB, SDVOSB, WOSB, and HUBZone entities. The contract is subject to the Davis-Bacon Act, Federal Contractor Minimum Wage requirements, and FAR standards regarding work hours, safety, and the Buy American Act.
Management & Training Corporation/Centennial Job Corps Center

POSTED

about 1 hour ago

DEADLINE

in about 1 month

AI Contract Overview

Show more

The Department of Veterans Affairs is conducting market research to identify qualified contractors for medical gas alarm cable services at the Marion VA Medical Center in Illinois. The project involves repairing and replacing deficient cabling and integrating CO2 alarms with the designated master alarm station to ensure reliable signaling. The selected contractor must provide all labor, materials, and testing necessary to complete the work within 45 calendar days of the notice to proceed. Interested vendors must submit a capability statement by August 20, 2026, demonstrating at least three years of experience in healthcare facility alarm systems. Key requirements include specific personnel certifications, such as ASSE 6040, and the ability to operate within an occupied medical environment. This notice is for planning purposes only and does not constitute a formal solicitation or a commitment to award a contract.

General Info

VA seeks contractors for medical gas alarm cable repairs at Marion VA Medical Center.

NAICS

238210 - Electrical Contractors and Other Wiring Installation Contractors

Place of Performance

Department of Veterans Affairs Marion VA Medical Center, Marion, IL, 62959-1188, USA

Set-Aside

NONE

Documents

1

Sources Sought Notice 36C25526Q0611 Medical Gas Alarm Cable Services

DOCX, Low prioritysources-sought
Low

AI Contract Breakdown

Uniform Contract Format

Sign up to view the full breakdown with detailed analysis of each section.

Timeline

PhaseClosed
Posted

Sources Sought

Response Deadline

Deadline has passed

Submission Closed

Find active opportunities like this

Start your free trial to discover similar active contracts, track opportunities, and build proposals with AI assistance.

Organization & Contact Information

Show more
AgencyDepartment Of Veterans Affairs → 255-NETWORK Contract Office 15 (36C255)
Contacts1 person available
OfficeLEAVENWORTH, KS, 66048, USA
Office AddressLEAVENWORTH, KS, 66048, USA
Contacts
Erika KobulnickyContract Specialist

Full Description

Show more
SOURCES SOUGHT NOTICE (Market Research Only) Department of Veterans Affairs (VA) NCO 15 Marion VA Medical Center (VAMC), Marion, IL Medical Gas Alarm Cable Services THIS IS NOT A SOLICITATION. This Sources Sought Notice is issued solely for market research and planning purposes and does not constitute a solicitation or an obligation on the part of the Government to acquire any products or services. Responses will be treated as information only. The Government will not reimburse any costs associated with responding. Information received may be used by the Department of Veterans Affairs to inform the acquisition approach and performance specifications. Submission of pricing, capabilities, and other market data is permitted in accordance with FAR 15.201(e). 1. PURPOSE The Department of Veterans Affairs (VA), NCO 15, is conducting market research in accordance with FAR Part 10 to identify capable and interested sources for services to repair and replace deficient medical-gas alarm cabling and integrate identified CO2 alarm(s) with the designated master alarm station at the Marion VAMC. Responses will help the Government finalize the Statement of Work (SOW), determine the acquisition strategy (including small-business considerations), and assess industry capability. The Government may not answer vendor questions at this time. 2. NAICS & BUSINESS INFORMATION NAICS Code: 238210 Electrical Contractors and Other Wiring Installation Contractors SBA Size Standard: $19.0 million (revenue). Socio economic status: Vendors shall indicate SDVOSB/VOSB (VetCert if applicable), WOSB, 8(a), HUBZone, Small/Large, etc. 3. PLACE OF PERFORMANCE Marion VAMC 2401 West Main Street Marion, Illinois 4. REQUIREMENT SUMMARY The VA Heartland Network 15 Contracting Office seeks a qualified contractor to furnish all labor, supervision, tools, equipment, materials, transportation, testing, documentation, and incidentals necessary to restore reliable, accurate, and identifiable alarm signaling for medical gas alarm circuits at the Marion VA Medical Center. Refer to the Statement of Work (SOW) which will follow this notice for additional details. 5. REQUESTED VENDOR INFORMATION Submit a capability statement (maximum 5 pages) including: Company Name, Address, and Point of Contact (name, phone, email). UEI, CAGE, and Tax ID. Business size under NAICS 238210 and socio economic status (SDVOSB, VOSB, WOSB, 8(a), HUBZone, Small/Large). Relevant experience performing medical gas alarm repair/installation/integration in occupied healthcare facilities (include references). Personnel qualifications/certifications (ASSE 6040/6010/6020/6030 as applicable) and any manufacturer training/authorization. Approach to cable repair/replacement, CO2 alarm integration, testing/commissioning, and measures to protect facility operations. Applicable contract vehicles (FSS, GSA, IDIQ, BPA) and contract numbers, if available. 6. SUBMISSION INSTRUCTIONS Responses Due: Thursday, August 20, 2026 by 7:00 AM Central Time (CT) Submission Method: Email PDF capability statement to: Erika M. Kobulnicky, Contract Specialist Erika.Kobulnicky@va.gov Subject Line: Sources Sought Response Medical Gas Alarm Cable Repair (Marion VAMC) 7. ADDITIONAL INFORMATION This notice shall not be construed as a commitment by the Government to issue a solicitation or award a contract, nor does it restrict the Government s acquisition approach. Based on responses, the Government may consider set asides for SDVOSBs, VOSBs, or another small business category; if inadequate small business response is received, alternate set asides or full and open competition may be considered in accordance with Public Law 109 461, Section 8127 (Veterans Benefits Act). 8. IMPORTANT NOTES This notice is for market research only and does not constitute a solicitation. The Government may not respond to questions at this time; vendors should submit capability statements based on this notice. Proprietary information must be clearly marked. No telephone inquiries will be accepted. The Statement of Work (SOW) will follow this notice and is subject to change based on market research and internal review. 9. STATEMENT OF WORK (TO FOLLOW) The SOW may be revised based on market research feedback. Statement of Work Medical Gas Alarm Cable Services 1. Purpose The Department of Veterans Affairs (VA), Marion VA Medical Center (VAMC), requires repair and replacement of deficient medical-gas alarm cabling and integration of identified CO2 alarm(s) with the designated master alarm station. The work is required to restore reliable, accurate, and identifiable alarm signaling for the affected medical-gas alarm circuits. 2. Scope and Required Outcome The contractor shall furnish all management, labor, supervision, tools, equipment, materials, transportation, testing, documentation, and incidentals necessary to complete the work described in this SOW, except for Government-furnished information expressly identified in Section 6. The completed work shall provide the following outcomes: Correct and reliable alarm communication between the identified originating locations and the designated master alarm station. Correct alarm identification, alarm activation, normal condition, and restoration indication for every affected circuit. Cable routing, termination, labeling, testing, and documentation that are compatible with the existing medical-gas alarm system, applicable codes and standards, manufacturer instructions, and VA facility requirements. A safe, operational, tested, documented, and VA-accepted system no later than 45 calendar days after the notice to proceed (NTP). 3. Locations of Work Work will occur at the Marion VAMC, 2401 West Main Street, Marion, Illinois, including the following known locations: Building 2, electrical room G115, outside the Kitchen. Building 42, mechanical room A109. Building 2, Room 245. Building 42, Emergency Department Medical Service Area (ED MSA) office/master alarm-panel location. Before starting work, the contractor shall conduct a site survey with the VA-designated representative to verify access, cable pathways, alarm-panel locations, cable endpoints, circuit identification, and existing conditions. The contractor shall promptly notify the COR in writing of site conditions that conflict with this SOW or prevent compliant performance. 4. Technical Requirements 4.1 Existing cable repair/replacement The contractor shall: Identify the affected existing medical-gas alarm cable connections and correct compromised, spliced, extended, improperly routed, mixed-up, or otherwise deficient connections within the defined work area. Furnish and install one 16-pair cable assembly, or other cable configuration approved in writing by the COR after field verification, between electrical room G115 in Building 2 and mechanical room A109 in Building 42. Verify the cable type, conductor size, shielding, listing, pathway, and terminations required for compatibility with the existing alarm equipment before procuring material. Materials shall be new and compatible with the existing system and manufacturer requirements. Route cable in an approved pathway and protect it from physical damage. The contractor shall not use unapproved pathways or penetrate fire/smoke barriers without VA approval and required firestopping. Terminate, identify, and label cables at both endpoints. Labels shall correspond to the approved cable schedule and affected alarm point(s).
4.2 CO2 master alarm connection The contractor shall: Furnish and install cable between Building 2, Room 245, and the designated master panel in the Building 42 ED MSA office. Verify the precise endpoint, signal type, alarm-panel compatibility, and required cable characteristics during the pre-work survey. Terminate, identify, and label the cable at both endpoints. Connect and commission the identified CO2 alarm(s) so that the required alarm condition, normal condition, and restoration indication are accurately displayed at the designated master panel.
4.3 Work plan and outages At least 10 calendar days after NTP, and before beginning field work, the contractor shall submit a work plan to the COR for review. The plan shall include the proposed schedule, site access requirements, cable routes, affected equipment/circuits, anticipated shutdowns or alarm impairments, protective measures, testing sequence, and contingency/restoration procedures. The contractor shall not begin work that may impair an alarm circuit, medical-gas alarm indication, or related facility operation until the VA-designated representative has approved the work window and required coordination. The contractor shall immediately notify the COR and VA-designated representative of an unintended alarm impairment, malfunction, or unsafe condition and shall take prompt action to restore the affected system.
5. Standards, Safety, and Contractor Qualifications 5.1 Applicable Requirements The contractor shall comply with all applicable Federal, state, local, VA, facility, manufacturer, and code requirements, including the edition of NFPA 99 adopted or otherwise specified by the VA facility for this project. The contractor shall comply with all VA facility requirements for medical-gas alarm impairments, permit-to-work procedures, infection-control measures, life-safety, firestopping, access, safety, and emergency restoration.
5.2 Contractor Experience The contractor shall demonstrate a minimum of three years of successful experience performing medical-gas alarm repair, installation, troubleshooting, testing, or commissioning in occupied healthcare facilities. Before beginning work, the contractor shall submit three references for completed projects of similar scope and complexity. Each reference shall include: Facility name and location. Description of the medical-gas alarm work performed. Contract or project completion date. Customer point of contact, telephone number, and email address. Identification of whether the work included alarm cabling, alarm-panel integration, troubleshooting, testing, commissioning, or system restoration.
5.3 Personnel Qualifications The contractor shall provide qualified personnel with demonstrated experience in medical-gas alarm systems, low-voltage control wiring, cable termination, alarm-panel troubleshooting, functional testing, and commissioning. Before beginning field work, the contractor shall submit the following for the project manager and each technician assigned to perform work: Resume or work history showing relevant medical-gas alarm and healthcare-facility experience. Applicable state or local electrical license, if required by the jurisdiction for the work being performed. Current training certificates, professional credentials, and medical-gas qualifications applicable to the assigned work. Documentation of manufacturer training, manufacturer authorization, or equivalent demonstrated competence for the existing medical-gas alarm equipment to be serviced.
Technicians performing maintenance, troubleshooting, testing, repair, or commissioning of medical-gas alarm equipment shall possess ASSE 6040 Medical Gas Systems Maintenance Personnel certification, or alternate written qualifications accepted by the VA medical-gas Responsible Facility Authority (RFA) before field work begins.
If the scope includes installation or modification of medical-gas-system components beyond alarm cabling and panel connections, including piping, outlets, valves, sensors connected to the pipeline system, or source equipment, personnel performing that installation work shall meet ASSE 6010 Medical Gas Systems Installer requirements.
If the VA RFA determines that inspection or independent verification is required, inspection and/or verification shall be performed by qualified ASSE 6020 Medical Gas Systems Inspector and/or ASSE 6030 Medical Gas Systems Verifier personnel, as applicable. The verifier shall be independent of the contractor performing the installation or repair work.
5.4 Medical-Gas Coordination Before beginning any work that may affect alarm function, medical-gas alarm indications, or facility operations, the contractor shall coordinate with the COR and the VA-designated Engineering Service and medical-gas RFA representative. The contractor shall submit a written work plan for VA review before beginning field work. At a minimum, the work plan shall identify: Affected alarm panels, circuits, alarm points, and cable routes. Proposed work sequence and work schedule. Planned alarm impairments, shutdowns, or temporary monitoring measures. Measures to protect patient safety and maintain facility operations. Testing and restoration procedures. Emergency contacts and contingency actions for an unplanned alarm-system malfunction or impairment. The contractor shall not impair, disconnect, alter, or place out of service an alarm circuit without prior coordination and approval through the VA facility s applicable impairment and permit-to-work process.
5.5 Safety and Protection of Government Property The contractor shall secure tools and equipment at all times. Tools and equipment shall be removed from open areas or placed in an approved secured location during breaks, at the end of each workday, and whenever not in active use. The contractor shall remove debris from the facility at the end of each workday and upon completion of the work. The contractor shall protect Government property and repair, at no additional cost to the Government and subject to VA approval, damage caused by the contractor or its employees. Smoking is prohibited on VAMC property.
6. Government-Furnished Information and Coordination The Government will provide reasonable access to the work locations and available existing system information, including available drawings, panel information, and known circuit information. The Government will identify the COR and the appropriate Engineering/medical-gas representative for coordination. The contractor shall not rely on incomplete or unavailable drawings as a basis to omit necessary site verification. The contractor shall document field-verified conditions and proposed deviations from the approved plan for COR review before implementing a deviation.
7. Submittals and Deliverables The contractor shall provide the following to the COR: Deliverable Due date Personnel qualifications/certifications, as applicable Before beginning field work Site-survey findings, cable schedule, and proposed routing Before installation Work plan, schedule, and outage/impairment plan Within 10 calendar days after NTP and before field work Product data for proposed cable, terminations, and related materials Before installation Daily status report while an alarm circuit is impaired, if applicable By the end of each affected workday Signed functional test report for every affected circuit/alarm point At completion of testing Marked-up as-built drawing and final cable schedule identifying routes, endpoints, terminations, labels, and approved deviations Before final acceptance Warranty information and closeout package Before final acceptance
8. Testing, Inspection, and Acceptance The contractor shall coordinate all testing with the COR and VA-designated Engineering/medical-gas representative. The contractor shall provide all test equipment and personnel required for contractor testing. For each affected circuit and alarm point, the contractor shall test and document, at a minimum: Continuity and correct point-to-point termination. Correct alarm activation and identification at the designated master alarm station. Normal-condition indication after testing. Correct restoration indication following simulated alarm activation. Proper labeling and correlation of field labels, panel labels, cable schedule, and as-built documentation.
The work is acceptable only when all of the following are complete: Each affected alarm point operates correctly and communicates the correct status and identification to the designated master panel. All cable routing, terminations, labels, and protective measures comply with the approved work plan, applicable requirements, and manufacturer instructions. Required test records and closeout deliverables have been submitted and are complete. VA has completed final inspection and provided written acceptance through the COR or other authorized VA official. The contractor shall correct deficiencies identified during testing or inspection at no additional cost to the Government and retest the corrected work.
9. Period and Place of Performance The period of performance is 45 calendar days after NTP. The place of performance is the Marion VAMC, 2401 West Main Street, Marion, Illinois, at the locations identified in Section 3 and any approved associated access pathways within the facility.
10. Payment Payment terms and contract line item structure will be established in the solicitation and resulting contract. Final payment is contingent on completion and Government acceptance of all required work and deliverables.
11. Contract Administration The COR may monitor performance and inspect deliverables but may not authorize changes to contract scope, price, period of performance, or other contract terms. Only the Contracting Officer may make such changes in writing. All contract administration functions will be retained by the Department of Veterans Affairs. The Contracting Officer will be the only person authorized to approve changes or modify any of the requirements under this contract. The Contractor shall communicate with the Contracting Officer on all matters pertaining to contract administration. Only the Contracting Officer will be authorized to make commitments or issue changes that affect price, quantity, or quality of performance of this contract. In the event the Contractor effects any such change at the direction of any person other than the Contracting Officer, the change shall be considered unauthorized and no adjustment will be made in the contract price to cover any increase in costs incurred as a result thereof.
12. Contracting Officer s Representative Pursuant to VAAR Provision 852.270-1, Representatives of Contracting Officer, the designated representative for this contract will be determined at award.
13. Acronyms and Definitions CDR: Contract Discrepancy Report. Report issued by the Government to the contractor to document a supply or service found to be unacceptable during contract performance. CLIN: Contract Line-Item Number. Unit of work (or service) to be performed (or delivered) by the contractor as a pay item. CO: Contracting Officer. A person with the authority to enter into, administer, and/or terminate contracts and make related determinations and findings. COR: Contracting Officer s Representative. An individual, including a Contracting Officer s Representative (COR), designated and authorized in writing by the contracting officer to perform specific technical or administrative functions. QA: Quality Assurance. Actions taken by the government to assure contracted services meet PWS requirements.
14. Personnel Non-Personal Services This is a non-personal services contract. Personnel rendering services under this contract are not subject either by the contract s terms or by the manner of its administration, to the supervision and control usually prevailing in relationships between the government and its employees. The Government shall not exercise any supervision or control over the contract service s performing services herein. Such contract service s shall be accountable solely to the Contractor who, in turn, is responsible to the Government. Contractor Compliance with the Immigration and Nationality Act of 1952 The Contractor shall comply with any and all legal provisions contained in the Immigration and Nationality Act of 1952, As Amended; its related laws and regulations that are enforced by Homeland Security, Immigration and Customs Enforcement and the U.S Department of Labor as these may relate to non-immigrant foreign nationals working under contract or subcontract for the Contractor while providing services to Department of Veterans Affairs patient referrals.
15. Hours of Operations Business hours for the VAMC follow: a. Administrative hours: Monday through Friday, 8:00 a.m. - 4:30 p.m. b. National Holidays: The 11 holidays observed by the Federal Government are: New Year s Day; Martin Luther King s Birthday; Presidents Day; Memorial Day; Juneteenth Independence Day Independence Day; Labor Day; Columbus Day; Veterans Day; Thanksgiving; Christmas; and any other day specifically declared by the President of the United States to be a national holiday. c. Off-Duty hours: Friday through Monday, 4:30 p.m. - 8:00 a.m.; Monday through Thursday, 4:30 pm 8:00 am.
16. Quality Control CONTRACT PERFORMANCE MONITORING. The COR may perform surveillance of services by any of the methods listed below: a. Observing actual performance; b. Inspecting the services to determine whether or not the performance meets the performance standards; and c. Review of any other appropriate records.
(1) When unacceptable performance occurs, the COR shall inform the Contractor and the Contracting Officer. This will normally be in writing unless circumstances necessitate verbal communication. In any case the COR shall document the discussion and place it in the COR file. When the COR determines formal written communication is required, the COR shall prepare a Contract Discrepancy Report (CDR) and present it to the contractor program manager. The Contractor shall acknowledge receipt of the CDR in writing. The CDR will specify if the Contractor is required to prepare a corrective action plan to document how the Contractor shall correct the unacceptable performance and avoid a recurrence. The CDR will also state how long after receipt the Contractor has to present this corrective action plan to the COR. The Government shall review the Contractor corrective action plan to determine acceptability. Any CDRs may become a part of the supporting documentation for any contractual action deemed necessary by the Contracting Officer.
(2) Pursuant to VAAR Provision 852.270-1, Representatives of Contracting Officer representative(s); The Government shall periodically evaluate the Contractor performance by appointing a Contracting Officers Representative (COR) to monitor performance to ensure services are received. The Government representative(s) shall evaluate the Contractor performance through inspections of observations, inspection of services or any other form of documentation and all complaints from VA personnel. The Government may inspect as each task is completed or increase the number of quality assurance inspections if deemed appropriate because of repeated failures or because of repeated customer complaints. Likewise, the Government may decrease the number of quality assurance inspections if performance dictates. The Government Contracting Officer shall make final determination of the validity of customer complaint(s).
(3) If any of the services do not conform to contract requirements, the Government may require the Contractor to perform the services again in conformity with contract requirements, at no increase in contract amount. When the defects in services cannot be corrected by re-performance, the Government may require the Contractor to take necessary action to ensure that future performance conforms to contract requirements at no additional cost to the Government.
17. Contract Security The C&A requirements do not apply and a Security Accreditation Package is not required.

More opportunities from Department Of Veterans Affairs → 255-NETWORK Contract Office 15 (36C255)

Same awarding agency

NAICS: 541330
New
Federal
C1AZ--657A4-26-125 Conduct Roof Deck Assessment Building 8 ME FY27 NRM
Solicitation # 36C25526R0183
The Department of Veterans Affairs Network Contracting Office 15 is requesting Standard Form 330 qualifications packages for architectural and engineering services to conduct a roof deck assessment at Building 8 of the John J. Pershing VA Medical Center in Poplar Bluff, Missouri. This procurement is 100 percent set-aside for Service-Disabled Veteran-Owned Small Business firms and is being conducted in accordance with the Brooks Act and FAR Subpart 36. The objective of the assessment is to evaluate the deterioration of the concrete roof deck and supporting welded steel bar joists, identify structural deficiencies and water infiltration, and recommend immediate life-safety mitigations and long-term repairs. The anticipated contract will be a Firm Fixed Price award with a design completion period of approximately 100 calendar days. The scope of work includes visual inspections, concrete sounding, non-destructive reinforcement scanning using Ground Penetrating Radar or electromagnetic locators, and an integrity assessment of steel bar joists and masonry parapets. Deliverables range from preliminary layouts and design development submissions to final bid documents provided in PDF, Auto-CAD 2025, and Revit formats, as well as final as-built drawings. Selection will be based on professional qualifications, specialized experience in active healthcare facilities, capacity to accomplish the work, past performance, and estimating effectiveness. Interested firms must submit their electronic SF 330 package, including a design quality management plan and a signed certification of compliance regarding subcontracting limitations, by October 28, 2026, at 2:00 PM Central Time.
Engineering Services

POSTED

about 5 hours ago

DEADLINE

in 14 days
View Details
NAICS: 811210
New
Federal
J061--Uninterruptible Power Supply St. Louis VA Medical Center
Solicitation # 36C25526Q0678
Solicitation 36C25526Q0678 is a firm-fixed-price, Indefinite Delivery Indefinite Quantity (IDIQ) contract issued by the Department of Veterans Affairs, Network Contracting Office 15. The contract requires comprehensive preventive maintenance, annual inspections, emergency services, and battery replacements for Uninterruptible Power Supply (UPS) units and batteries at the John Cochran and Jefferson Barracks VA Medical Centers in St. Louis, Missouri. The period of performance is five years, with a contract ceiling of 215,000 dollars and a guaranteed minimum of 43,000 dollars. While initially a small business set-aside, Amendment 0003 transitioned the solicitation to unrestricted competition. Award will be based on the Lowest Price Technically Acceptable (LPTA) process, where quotes are rated as either technically acceptable or unacceptable based on their ability to meet the Performance Work Statement and specifications. To be considered responsive, offerors must provide a completed price schedule, evidence of relevant past performance, and Original Equipment Manufacturer (OEM) authorization letters for each OEM represented in the PWS. Technicians must be certified in OSHA 1910 Subpart S and NFPA 70E safety standards. Following several amendments, the final quote submission deadline is October 28, 2026, with an optional site visit scheduled for October 22, 2026.
Electronic and Precision Equipment Repair and Maintenance

POSTED

1 day ago

DEADLINE

in 7 days
View Details
NAICS: 541519
New
Federal
NEXODOSE DOSE MONITOR
Solicitation # 36C25527Q0017
The Department of Veterans Affairs Network Contracting Office 15 is conducting market research through a Sources Sought announcement to identify qualified businesses capable of providing software maintenance, support, and interface services for the DoseMonitor/Nexodose radiation dose tracking platform. This requirement supports clinical safety and regulatory compliance for VA medical facilities across Kansas and Missouri, including sites in Topeka, Leavenworth, Kansas City, St. Louis, Columbia, and Poplar Bluff. The scope of work includes annual maintenance and support for the DoseMonitor/Nexodose application and outbound API/HL7 interfaces, as well as the one-time installation and ongoing support of bi-directional HL7 interfaces with the Oracle Health/Cerner electronic health record system in both production and test environments. The anticipated period of performance consists of a base year from February 1, 2027, to January 31, 2028, with one option year extending to January 31, 2029. Work will be performed primarily remotely, though on-site support may be required for installation and troubleshooting. Contractors must adhere to VA Handbook 6500, HIPAA privacy rules, and HL7 v2.x messaging standards. Interested firms must be registered in SAM.gov and provide a capability statement, product brochures, and an authorized service provider letter from the OEM. Responses should be submitted to Betty Flores at veronica.flores@va.gov by October 29, 2026.
Other Computer Related Services

POSTED

2 days ago

DEADLINE

in 8 days
View Details
NAICS: 236220
New
Federal
Z1DA--657-24-111JC FCA-Renovate Ambassador Suite & Main Lobby B1
Solicitation # 36C25526R0167
Solicitation 36C25526R0167 is a Firm Fixed Price construction project issued by the Department of Veterans Affairs for the renovation of the Ambassador Suite and Main Lobby in Building 1 at the John Cochran VA Medical Center in St. Louis, Missouri. The project scope covers approximately 3,562 square feet and includes selective demolition, improving ADA access, repairing skylight leaks, updating finishes, correcting HVAC and electrical deficiencies, and replacing a revolving door with sliding or swinging doors. The estimated contract value ranges between $1,000,000 and $5,000,000, with a performance period of 330 calendar days. This is a 100% Service-Disabled Veteran-Owned Small Business (SDVOSB) set-aside under NAICS code 236220. Award will be based on the Lowest Price Technically Acceptable (LPTA) selection process. To be considered technically acceptable, offerors must demonstrate relevant past performance in hospital construction, provide a qualified site superintendent and safety officer with at least five years of experience, and submit a comprehensive safety plan for working in an active healthcare facility. Compliance with the Davis-Bacon Act and specific subcontracting limitations is required, mandating that no more than 85% of government payments be made to non-SDVOSB or non-VOSB firms. Proposals are due by November 4, 2026.
Commercial and Institutional Building Construction

POSTED

5 days ago

DEADLINE

in 14 days
View Details

Find Active Opportunities Like This

Get AI-powered intelligence on the opportunities still open

Every page of the solicitation package shredded into a compliance breakdown

AI-powered matching based on your capabilities and past performance

Competitor and incumbent history on the requirement

Automated alerts on amendments, Q&A deadlines, and award

Miguel
Hillary
Keith Deutsch
Christine

Join 750+ contractors already using CLEATUS