This Combined Synopsis/Solicitation opportunity from Department Of Veterans Affairs was posted on June 23, 2026. The submission period has ended. Browse the details below for market research, or find similar active opportunities.
N063--QUESTIONS & ANSWERS. NURSE CALL SYSTEM FOR DC VAMC. "BRAND NAME OR EQUAL TO IAW SOW". OEM LETTER FROM THE MANUFACTURER IS REQUIRED. IF NO OEM LETTER VENDOR WILL BE CONSIDERED NON-RESPONSIVE.
Contract Overview
Solicitation details, issuing organization, response deadlines, documents, and interested companies for this government contract opportunity.
Active Opportunities Like This One
AI Contract Overview
The contract solicitation 36C24526Q0598 for a Nurse Call System Upgrade at the Washington DC VA Medical Center requires a complete replacement and expansion of the existing system in the 3DE Mental Health Unit, which is currently partially non-functional due to water damage affecting 14 call buttons across three bathrooms. The scope includes installing 19 new push-button call devices, three localized alarm panels, and four dome lights, with the expansion adding a fourth bathroom and six additional call buttons requiring new rough-ins and cable runs. The system must operate as a standalone solution with no integration to the facility’s existing nurse call infrastructure, and all new equipment must be installed in conduit per VA regulations, regardless of whether existing or new pathways are used. All existing hardware, including the outdated Rauland 4 nurse call system, must be removed by the Contractor, and the new system must include high and low mounted push buttons to accommodate patient accessibility in behavioral health settings. The Contractor is responsible for determining the location of the head end equipment, existing panel placements, and all site-specific conditions, as no site plans or markups of existing infrastructure are provided by the Government. The installation must comply with an ICRA Level 2 standard with potential escalation to Level 4, necessitating stringent infection control measures, and must adhere to VA TIL Section 27-52-23 for critical life safety systems. The procurement is a total small business set-aside under NAICS code 334290 with an 800-employee size standard, and only offers from certified small businesses will be considered. A manufacturer’s OEM letter of authorization is mandatory; failure to provide it will result in non-responsiveness. The system must meet "brand name or equal" requirements as defined in the Statement of Work and must be procured through an approved manufacturer distribution channel. The Offeror must submit a completed Buy American Certificate for line item 1, and while the existing system may not have complied with Buy American provisions, the new equipment must. Warranty obligations include a one-year manufacturer parts warranty and a three-year workmanship warranty covering both new installations and any retained or integrated existing components. The entire project must be completed and fully operational within 90 days of contract award, with delivery to the Washington DC VAMC warehouse following FOB Destination terms. Evaluation is based on technical compliance, past performance, delivery lead time, Buy American status, and total price, with technical compliance as the primary factor.
General Info
Agency
Contract Value
$81,559.68NAICS
Place of Performance
Washington DC VA Medical Center 50 Irving Street NW, MDSet-Aside
Awardee
Award Issued Date
Timeline
Submission Closed
Organization & Contact Information
Full Description
ANSWER: ICRA level determination is dependent on the extent of work required, approval by Infection Control Specialist and is the responsibility of the Contractor to meet. Per Contract Specifications, work is likely to be level 2 but has the potential to go up to level 4.
Attachment 1 RFQ 36C24526Q0598, Page 19, Section 7. Period of Performance states Delivery within 90 days of contract award. Is this 90-day period of performance for the entire scope of the project or just for delivery of the requested hardware? ANSWER: The 90-day period of performance is for the entire scope of the project. Does the government require that behavioral health devices are true stainless-steel flush mounted anti-ligature devices, not generic devices with plastic coverings overtop? ANSWER: Beyond specifications of Contract. According to the DCVA staff, the current system has been inoperable for more than five (5) years. Has the government considered replacing all of the existing and new areas with a new system that will remain in support by the manufacturer longer and provide more features/capabilities? ANSWER: The government has considered replacement of the existing system. In most cases the government hosts a site visit for interested bidders, especially when construction tasks i.e. conduit, cabling, paint and patch, ICRA, etc. are required. Will the government allow for a site visit for this opportunity to allow bidders to better understand the scope of work required and current system conditions? ANSWER: Contracting requirement so I will defer but facility can allow a site visit after award without extension of POP. ANSWER: No site visit during RFQ. Please submit your response IAW SOW. Attachment 3 SOW, Page 1, 3.0 states The expansion will include a 4th bathroom along with six additional calls buttons that will require rough-ins and cable runs to the panel. There is no indication or icon in the drawings for a panel, where is the existing panel? ANSWER: There is no mark-up available for existing panels and would be the responsibility to be determined by the Contractor. Attachment 3 SOW, Page 2, has a quantity of 3 for dome lights but page 1 under locations as shown indicates 4 dome lights. Will the government clarify which is correct? ANSWER: There are 3 existing dome lights, but the Contract requirement is for a total of 4 dome lights. Will the government confirm the location of head end equipment and if power supplies or UPSs will be required in the event of a power outage? ANSWER: There is no mark-up available for existing equipment and would be the responsibility to be determined by the Contractor. UPS is not included in the requirements of this Contract. Attachment 3 SOW, Page 2 states that A manufacturer parts warranty for a minimum of 1 year shall be included. This is usually a standard five (5) year warranty on hardware required in VA TIL Section 27-52-23. Will the government require a five (5) year manufacturer s hardware warranty? ANSWER: The warranty requirement is as written in the SOW. Attachment 3 SOW, Page 2 states that A workmanship warranty for a minimum of 3 years shall be included. Does this include the existing system or only net new devices and associated installation work? ANSWER: Warranty requirement applies to the final deliverable regardless of whether existing or new devices are utilized.
NOTE: VHA-SERVICES-RFQ: 36C24526Q0598 VHA-SERVICES-RFQ-AMEND 36C24526Q0598 - 0001
All quotations shall be sent via email to Mohsin Abbas, Contract Specialist, Mohsin.Abbas2 @va.gov . Submission shall be received no later than 3:00 PM EST on June 30, 2026. Late submissions shall not be accepted. FAR 52.212-1(f)(1), RFO 52.212-1(c)(1). Only e-mailed quotes will be accepted. It is the responsibility of the vendor to ensure the Government s timely receipt of their quote.
Any questions or concerns regarding this solicitation should be forwarded in writing via e-mail to the Contract Specialist, Mohsin Abbas, Mohsin.Abbas2@va.gov.
More opportunities from Department Of Veterans Affairs → 245-NETWORK Contract Office 5 (36C245)
Same awarding agency
Find Active Opportunities Like This
Get AI-powered intelligence on the opportunities still open
Every page of the solicitation package shredded into a compliance breakdown
AI-powered matching based on your capabilities and past performance
Competitor and incumbent history on the requirement
Automated alerts on amendments, Q&A deadlines, and award
Join 650+ contractors already using CLEATUS