W065--Lease SPS Reverse Osmosis - Brooklyn & Manhattan

Awarded
36C24226Q0560Federal

Contract Overview

Solicitation details, issuing organization, response deadlines, documents, and interested companies for this government contract opportunity.

AI Contract Overview

Show more

The Department of Veterans Affairs, through the 242-NETWORK Contract Office 02, has issued a request for quotation under solicitation number 36C24226Q0560 for the lease and maintenance of reverse osmosis critical water systems at the Brooklyn and Manhattan VA Medical Centers, classified under NAICS code 221310 as a total small business set-aside. The contract is envisioned as a firm-fixed-price award with a one-year base period from July 1, 2026, to June 30, 2027, and four optional one-year extension periods through June 30, 2031, allowing for a potential five-year performance duration. The contractor must provide not only the leasing of SPS reverse osmosis equipment but also ongoing maintenance, repair, inspection, testing, and monthly sanitization of water loops and tanks, alongside comprehensive water quality sampling and testing at least once per month for both critical and utility water. All services must adhere to ANSI/AAMI ST 108:2023 standards and include documentation of equipment specifications, technician credentials, and validation protocols to confirm system functionality and water quality compliance. Acceptance of services is governed by performance-based criteria established by the vendor, with a 60-day presumption of acceptance unless previously rejected, and immediate corrective action required within 24 hours if water quality falls outside established parameters at no additional cost to the government. The contract incorporates a suite of FAR and VAAR clauses, including 52.212-4 and 52.212-5 as foundational terms, with deviations such as 52.219-8 for small business utilization, 52.222-90 addressing DEI discrimination, and 52.240-91 for security prohibitions, all linked as addenda to the base clause. The 52.217-8 and 52.217-9 options allow for service and term extensions, while 852.232-72 mandates electronic invoicing via the Tungsten Network portal in monthly arrears using EFT through SAM. Performance is restricted to the Brooklyn and Manhattan VAMC sites, with payment tied to adherence to strict accounting, insurance, and bonding requirements including general liability of $500,000 per occurrence, auto liability of $200,

General Info

Lease of redundant reverse osmosis water systems for VA, small businesses, ANSI/AAMI compliant, maintenance included.

Contract Value

$569,881.82

NAICS

221310 - Water Supply and Irrigation Systems

Place of Performance

NY, USA

Set-Aside

SBA

Award Issued Date

Documents

7

Amendment 0001 to Solicitation 36C24226Q0560 for Water Quality Testing Services

DOCX, High priority: read this firstamendment
High

Amendment 0002 to Solicitation 36C24226Q0560 - Extension and Q&A

DOCX, High priority: read this firstamendment
High

RFQ 36C24226Q0560 Lease SPS Reverse Osmosis Brooklyn & Manhattan VAMC

DOCX, High priority: read this firstrfq
High

Solicitation 36C24226Q0560 Lease and Maintenance of Reverse Osmosis Systems

DOCX, High priority: read this firstrfq
High

Sources Sought Notice 36C24226Q0560 for Reverse Osmosis System Lease at Brooklyn & Manhattan VAMC

DOCX, High priority: read this first11 pages · sources-sought
High

Wage Determination Register for New York - Service Contract Act

PDF, High priority: read this firstwage-determination
High

36C24226P0771 - Lease SPS Reverse Osmosis - Brooklyn & Manhattan

DOCX, Low priorityaward
Low

AI Contract Breakdown

Uniform Contract Format

Sign up to view the full breakdown with detailed analysis of each section.

Timeline

2 updates
PhaseAwarded
Posted

Sources Sought

Type Changed

Sources Sought → Presolicitation

Type Changed

Presolicitation → Solicitation

Notice updated

Response deadline changed from Jun 15, 2026 to Jun 18, 2026. The notice description changed.

Notice updated

Response deadline changed from Jun 18, 2026 to Jun 24, 2026. The notice description changed.

Awarded

Contract was awarded

Ready to pursue this opportunity?

Get full access to track this contract, build proposals with AI assistance, and manage your pipeline.

Organization & Contact Information

Show more
AgencyDepartment Of Veterans Affairs → 242-NETWORK Contract Office 02 (36C242)
Contacts1 person available
OfficeALBANY, NY, 12208, USA
Office AddressALBANY, NY, 12208, USA
Contacts
Dolores Sachuk-ThompsonContracting Officer

Full Description

Show more



?.## STATEMENT OF WORK
LEASE FOR STERILE PROCESSING REVERSE OSMOSIS (RO) CRITICAL WATER SYSTEM, MAINTENANCE AND WATER QUALITY TESTING BROOKLYN VAMC AND MANHATTAN VAMC
Introduction: The contractor shall provide, install and maintain, under lease arrangement, a Reverse Osmosis (RO) critical water treatment system in the Sterile Processing Service (SPS) located at the Brooklyn VA Medical Center and Manhattan VA Medical Center. The system shall have a built-in means of redundancy, such that the backup system can be operated should the primary system be out of service. Additionally, the contractor shall test utility water used in both the Brooklyn VAMC; 800 Poly Place, Brooklyn, NY 11209, SPS Department and Manhattan VAMC; 423 E. 23rd. St., New York, NY 10010, SPS Department as well as steam (as condensate), and the final rinse water of the Automated Endoscope Reprocessors. Special Conditions: The facility has an existing critical water system leased by the VA which will be removed once the contract elapses for the new contractor. During the removal process of the old system, the vendor must install and operate a temporary critical water system meeting all contract requirements, to include testing and monitoring, until vendor s permanent system can be installed. Coordinate removal/set up/maintenance so at least one (1) facility is running at all times. Privacy: All contractor employees and subcontractors participating under this contract are required to complete the VA's Privacy training requirement - VA Privacy Training for Personnel without Access to VA Computer Systems or Direct Access to or the Use of VA Sensitive Information Contractors must provide certifications of completion to the COR during each year of the contract. Training expires 365 days after the training is taken. This requirement is in addition to any other training that may be required of the contractor and subcontractor(s). The COR for the contract is responsible for ensuring the contractor takes the training and to follow up with the annual requirements. Critical Water System Requirements: The system must produce water meeting all requirements of the American National Standard Institute/Association for the Advancement of Medical Instrumentation (ANSI/AAMI) ST108:2023 Water for the processing of medical devices , including but not limited to:
At a minimum, the system shall incorporate the following technologies: appropriate pre-treatment including multi-media filter to remove suspended solids, water softener, carbon filter, reverse osmosis, ultraviolet light disinfection, deionization, and endotoxin filtration.
System shall have sufficient capacity to provide 30 gallons per minute of treated water and be equipped with a minimum 125-gallon treated water storage tank. The system shall provide recirculation at a rate of 3 to 5 feet per second as stipulated by ANSI/AAMI ST108:2023.
System dimensions shall not exceed 72 length, 36 depth and 72 height and shall fit within the designated room.
System shall provide acceptable quality Critical Water (as defined by ANSI/AAMI ST108:2023) to various outlets in the existing SPS suite including decontamination sinks, washer-disinfectors, ultrasonic equipment, spray nozzles and other devices designated. Systems shall be on a fully recirculating loop to allow for full and effective disinfection of the system piping and have a fully drainable tank for complete and faster disinfection.
The system shall have a properly sized final loop filter of 0.2 micron or tighter/smaller (absolute rated, not nominal rated), with endotoxin removal capabilities necessary to achieve endotoxin specification as defined in ANSI/AAMI ST108:2023 Table 2. This filter should be located at the point of connection to the loop supply and sized to meet or reasonably exceed the intended flow rate of the loop.
Vendor shall install sanitary water sampling ports at all critical locations including loop out after final filtration, right before re-entry into the storage tank, and before all individual points of use (e.g. immediately before the inlet to the washer-disinfectors, before inlet to the ultrasonics, etc.).
The system shall provide Critical Water meeting the quality requirements outlined in Table 2. This includes, but is not limited to, specified quality requirements for pH, Conductivity/Resistivity, Total Alkalinity, Total Hardness, Heterotrophic Plate Count (HPC) Bacteria, Endotoxin, Color and Turbidity, Total Organic Carbon (TOC), Endotoxin, and Ionic Contaminants (Aluminum, Chloride, Copper, Iron, Manganese, Nitrate, Phosphate, Sulfate, Silicate and Zinc).
The vendor shall provide delivery, uncrating, assembly, final electrical and plumbing connections and all required commissioning of the critical water system.
Critical alarms, such as those associated with deionizer exhaustion or low water levels in a storage tank, should be configured to be audible and/or visual both in the water treatment area and in the area where the water is used or monitored to alert operators to faults or failures of the supplied utilities. Alarms must be installed for the continuous monitoring of conductivity as well as low level tank alarms. Critical Water System Maintenance Requirements: The vendor shall conduct a training plan for all necessary VA personnel on the operation and basic troubleshooting of the new system to include any system assessments required by the end user per manufacturer s instructions.
The Vendor shall perform routine maintenance of the system at a minimum frequency of once per month and as outlined in the equipment manufacturer s instructions for use. Additional service visits shall be made at whatever frequency is needed to ensure that the system is operational and produces water at the specified rate and quality standards.
Monthly service visits shall include loop and tank sanitization using procedures stipulated by ANSI/AAMI ST108:2023. Additional service visits shall be made for loop disinfection and related services at whatever frequency is required to ensure that the system is operational and producing water at the specified rate and quality standards.
Post construction or extended shutdown, vendor shall perform all procedures as outlined in ANSI/AAMI ST108:2023 Section 13.1 Post construction or extended shutdown .
When the piping is replaced, or system modification is completed on the Critical Water system, the vendor shall disinfect the system piping and storage.
Disinfection shall be performed after installation, expansion, major repairs, non-compliant bacterial levels, or other changes to the system (e.g., nearby building construction or water supply interruptions).
Emergency Service Visits: the vendor shall respond on-site to all emergency service calls within a maximum of 4 hours.
Critical Water System Installation and Operation Qualification Requirements: The installation and operation qualification requirements are specific to the treatment methods selected. The vendor shall establish the acceptance criteria for quality characteristics of the water and operating parameters of the water treatment system.
The vendor shall define the water treatment system installation acceptance criteria and document that the installation meets those criteria.
The vendor shall perform protocols that verify that the water treatment equipment and system alerts and controls are functional. Critical Water System Performance Qualification: Determining the dependability of a water treatment system requires a period of monitoring and observation. After the installation and qualification of a water treatment system or when the system has been modified or undergone significant maintenance/repair, the vendor will perform a validation program consisting of intensive daily sampling and testing at post water treatment system test points and at the point-of-water-use and then performance qualification testing resumes at a reduced frequency for at least one calendar year to identify seasonal changes.
After installation or existing system modifications (any changes to the system that may affect either the operating conditions or otherwise disrupt the steady state of operation), the vendor will test for the water quality measurements shown in Table 2 for three (3) days. Ideally, vendor will test on three (3) consecutive days.
The vendor shall test for the water quality measurements according to the daily schedule in Table 3.
The vendor shall demonstrate that the system has achieved a stable operation when three consecutive samples all test within range.
When three (3) consecutive samples tests are all within range, the frequency of testing will be reduced to the frequencies outlined in Table 4.
During performance qualification, the entire water treatment system shall be assessed by the vendor. The number of sample sites for daily sampling may be reduced, provided that all sample sites are examined at some point during the performance qualification. When all sample sites are examined, routine monitoring can be reduced when approved by the facility, on the basis of the results of the performance qualification.
The following sample sites will be evaluated by the vendor during the performance qualification: The incoming water Following each treatment step; and At the point-of-water-use including: Critical Water at point-of-use or as close to the point at which the water will be representative of water that will contact the device. Critical Water System Routine Monitoring Requirements (After One Year): After the performance qualification period (one year), the vendor shall perform routine testing of the critical water system at the point of critical water generation as well as testing at the point of use. The locations and frequencies shall be completed according to Table 1. This table was developed from the AAMI requirements found in Table 5 and Table 6 from ANSI/AAMI ST108:2023.
The routine monitoring tests shall be based on the acceptance criteria defined in Table 2.
The recommendations for frequency of testing in Table 1 are the recommended minimum frequency. When problems or issues arise with the water quality, it is necessary to increase the frequency until they are resolved.
RO systems and Deionizers should be monitored using continuous reading monitors that measure conductivity (or total dissolved solids).
Analyses for chemical contaminants should be performed upon installation of the RO system, when membranes are replaced, and at least annually thereafter to ensure that established limits are met.
Other than pH analysis (which may be done via appropriate field instrumentation), all samples shall be analyzed by a New York State Department of Environmental Laboratory Accreditation Program (ELAP) certified laboratory with experience in microbial and chemical testing of potable water. The lab shall also possess American Industrial Hygiene Association (AIHA) accreditation under the Environmental Microbiological Laboratory Accreditation Program (EMLAP).
Bacterial tests must be performed via the Heterotrophic Plate Count (HPC) methodology as outlined in ANSI/AAMI ST108:2023. Note: in the event that bacterial results exceed the recommended threshold of 10 Colony Forming Units Per Milliliter (CFU/ml), the outlets exceeding the threshold should be evaluated for potential sources of contamination. The frequency of sampling should be increased to weekly, until the situation is resolved (as indicated by a minimum of two consecutive rounds of testing with acceptable results). Sampling Procedure: The vendor shall adhere to strict quality control procedures to ensure sample integrity, as outlined hereafter. The Contractor must furnish all sample collection equipment. The vendor should be escorted by facility water management personnel during all sample collections. Bottles must be sterile, appropriate for the analysis, properly preserved and promptly transported to the laboratory after sample collection under strict chain of custody procedures. When taking any test, the vendor should carefully follow the manufacturer s written IFU for accurate results.
The consultant performing sampling shall collect samples utilizing the nomenclature and from locations stipulated by the VA.
Samplers shall utilize dedicated sterile bacteriological sampling bottles provided by the certified laboratory. Bottles must have a plastic tamper evident seal to ensure that they are sterile. The seal shall only be removed immediately before sample collection. The volume of the sample container should be in the range of 100 to 300 milliliters.
The sampler shall don a clean pair of nitrile gloves prior to sampling. Sanitize the gloves with Purell or other alcohol-based hand sanitizer.
Samples shall always be collected before disinfection of the water treatment system. If repeat cultures are performed after the system has been chemically disinfected (e.g., with hydrogen peroxide, chlorine, or peracetic acid), the system should be flushed completely before samples are collected.
Locations to be sampled for bacterial analysis must be flushed for a minimum of five (5) minutes prior to sample collection.
Following the initial flushing period, immerse the distal point tap/faucet/outlet/sprayer in a solution of minimum 70% alcohol for at least five (5) minutes. Following the alcohol immersion, run the tap again for at least 30 seconds prior to sample collection.
The collection container should not be made of polypropylene.
When opening the sampling bottle for sample collection, take care to ensure that the sample contact area of the cap does not come in contact with any surface. The caps should be set down on a clean surface, topside against the clean surface.
Fill the bottle without allowing the bottle to come into contact with the faucet/distal point in question. Do not overflow.
Following sample collection, immediately seal the bottle with the cap. Ensure the bottle is labeled in accordance with stipulated sampling nomenclature as well as filling in any information required by the laboratory such as collection time.
As stipulated by ANSI/AAMI ST108:2023, lab analysis should begin within 8 hours of sample collection. By no means should more than 24 hours elapse between sample collection and analysis. Exceeding the 24-hour threshold will invalidate the results.
Immediately upon sample collection, samples shall be put on ice or refrigerated such that temperature is maintained between 2°C and 8°C (35.6°F and 46.4°F). Samples should be delivered to the laboratory as soon as feasible following collection. Automated Endoscope Reprocessor (AER) Water Testing: The vendor shall take a sample of the final rinse water from the automated endoscope reprocessors in accordance with the steps outlined in the equipment manufacturer s instructions for use. The sample shall be taken using aseptic technique.
The vendor shall test the AER s final rinse water per the locations and frequencies outlined in Table 1.
The routine monitoring tests shall be based on the critical water acceptance criteria defined in Table 2.
The recommendations for frequency of testing in Table 1 are the recommended minimum frequency. When problems or issues arise with the water quality, it is necessary to increase the frequency until they are resolved. Steam Quality Testing: The vendor shall test the steam condensate per frequency and sample locations outlined in Table 1.
The routine monitoring tests shall be based on the acceptance criteria defined in Table 2.
The recommendations for frequency of testing in Table 1 are the recommended minimum frequency. When problems or issues arise with the steam condensate quality, it is necessary to increase the frequency until they are resolved. Utility Water Testing: The vendor shall test the utility water per frequency and sample locations outlined in Table 1.
The routine monitoring tests shall be based on the acceptance criteria defined in Table 2.
The recommendations for frequency of testing in Table 1 are the recommended minimum frequency. When problems or issues arise with the steam condensate quality, it is necessary to increase the frequency until they are resolved.
Analytical Reporting: The consultant shall provide complete laboratory reports including VA stipulated sampling location nomenclature, results summary, chain of custody and original laboratory reports to the Engineering Chief, Safety Manager, SPS Chief and other designated VA representatives within 72 hours of completion.
Exceptions (Water Quality Issues/Exceedances of the AAMI guidelines) shall be promptly reported to the SPS Chief, Engineering Chief and Safety Manager. Response Actions for Water Quality Exceedances: When water quality leaving the vendor s treatment system is found to be out of range as defined in ANSI/AAMI ST108:2023, the vendor is responsible for promptly (within 24 hours) making all needed corrections at no cost to the VA. This may include but is not limited to flushing and resampling, filter changes, sanitization and other preventative measures. Following response actions, the vendor shall perform resampling to verify that acceptable water quality is being provided.
When microbial quality levels exceed specifications, a repetition of the disinfection and sampling should occur at the location to confirm the results. Upon confirmation of the exceedance, immediate action is required to reduce the microbial contamination. The distribution loop shall be disinfected using methods described in Annex E in ANSI/AAMI ST108:2023 to ensure the microbial contamination is not being introduced through the distribution system.
Automated instrument reprocessors (e.g., washer disinfectors, washer decontaminators, AERs) should be flushed with 1,000 ppm of a chlorine-releasing agent after the first instance of persistent unacceptable microbial levels is detected. Other agents that could be considered for flushing include chlorine dioxide, peracetic acid, hydrogen peroxide, and super oxidized water.
During the period between the receipt of the out or range results and the verification of restoration of compliant water quality, the vendor shall activate the redundant (backup) critical water system as referenced herein.
When water quality at outlet (distal) points is found to be out of range as defined in ANSI/AAMI ST108:2023, while water quality leaving the vendor s system is in range, the vendor shall make specific recommendations to the VA in order address the nonconformance. When the exceedance results from contamination in the VA s piping distribution system, the vendor shall advise the COR of the result and offer suggestions to the VA as to how to respond. The COR may request that the vendor perform additional distribution system sanitization (in addition to the monthly routine sanitization) and flush out (additional sanitization, if required, will be covered under the designated contract CLIN for Unscheduled Maintenance ). Upon completion of sanitization, the vendor will be requested to resample as stipulated in contract CLIN.
Hours of coverage: Will be from 8:00 am through 4:00 PM Monday through Friday, excluding Federal Holidays; all work done outside the normal coverage time should be coordinated through the COR. Preventive maintenance should be scheduled, during normal hours, with at least two weeks advance notice to the VA COR. Preventive maintenance will be done at the manufacturer s recommended frequency or greater (at a minimum, monthly).
Hardware and software upgrades will be scheduled and performed during normal working hours, at no additional charge to the VA. These upgrades are those required for the correct function of the system; not those purchased to enhance function.
The contractor will provide all hardware and software necessary for unscheduled repairs. The contractor s field service engineer will respond by phone within 2 hours and respond on site within 4 hours to correct the problem.
The Contractor shall be responsible for all routine maintenance and sampling and testing to ensure acceptable quality of the treated water supplied by the Reverse Osmosis system. The Contractor is responsible for all scheduled maintenance in accordance with manufacturers schedule for maintenance. All replacement parts for normal wear and tear items shall be covered by this contract and provided and installed at no additional charge to the VA.
The contractor must have trained personnel within a 90-mile radius of Brooklyn VA Medical Center and Manhattan VA Medical Center, New York, with an adequate spare parts inventory in some area office and proof of manufacturers or equivalent maintenance training of field service personnel. Each scheduled and unscheduled visit will result in the production of a written report, by the contractor s representative, in a format negotiated between the VA-COR and the contractor.
16.) Spare Parts:
Adequate supply of new spare parts shall be maintained by the Contractor to ensure prompt preventive maintenance and repair services.
If it becomes evident during a warranty period that a component(s) is not functioning properly in accordance with manufacturers or Contract requirements, or if in the opinion of COR, excessive maintenance and attention must be employed to keep the component(s) operational, then the component(s) shall be removed and a new component(s) meeting all requirements shall be installed.
17.) Safety:
Contractor shall adhere to ANSI/AAMI ST108:2023 standards, and other regulatory requirements. Safety shall be the Contractor s highest priority.
During contract performance, the Contractor shall follow Government s safety policy and standards for safe work practices and take such safety precautions as the VA Safety Officer or designee or the CO/COR may determine to be reasonably necessary to protect the lives and health of occupants of the building. The Contractor shall comply with VA smoking policy.
Patient, employee, visitor, and Contractor personnel safety shall always be maintained. The Contractor is responsible for the safety and health of his/her employees.
Contractor shall be notified of any non-compliance with safety regulations. The Contractor shall, after receipt of such notice, immediately correct the condition and notify the Contracting Officer in writing that the condition has been corrected. If, in the opinion of the CO/COR or the Government s Safety Officer, the condition is life threatening he/she shall instruct the Contractor to stop work until the condition has been corrected. If the Contractor fails or refuses to comply promptly, the Contracting Officer may issue an order halting all or any part of the work. The Contractor bears all costs of stop-work time and correcting safety hazards.
Contractor shall immediately report to the CO/COR in a manner and on the forms prescribed by the Government, accidents resulting in injury, trauma, death, hazardous exposures, and occupational disease involving an elevator or its maintenance.
Written reports for incidents shall be provided within five (5) working days and include the cause, what, where, and when the incident occurred as well as repairs and tests performed to correct the cause of the incident.
18.) Quality Control Requirements: Contractor shall establish a Quality Control Program to assure the requirements of the Contract are provided as specified. The Contractor shall submit this program to the CO/COR, for acceptance within ten (10) days of Notice to Proceed. 19.) Unscheduled Maintenance and Repair Actions:
Unless otherwise directed by the CO the Contractor shall not perform any work outside of the Scope of this Contract without authorization from the CO and the issuance of a separate Purchase Order. Failure to receive prior approval for out-of-scope work will most likely result in the denial of compensation. 20.) Period of Performance:
The period of performance shall be for one (1) Base Year of 12 months and Four (4) 12-month option years. The Period of Performance reads as follows:
Base Year - July 1, 2026 June 30, 2027 Option Year I July 1, 2027 June 30, 2028 Option Year II July 1, 2028 June 30, 2029 Option Year III July 1, 2029 June 30, 2030 Option Year IV July 1, 2030 June 30, 2031 TABLE 1: Water Quality Routine Testing Frequency for Critical Water, Utility Water, AER Final Rinse, and Steam Condensate for each (Brooklyn and Manhattan) of the Medical Centers will be:
Water Quality Measurement Source Sampling Site(s)
# of Samples per event Sampling Frequency # of Samples per year
BROOKLYN pH Critical Water After the last treatment step at the water generation system At the point the distribution loop enters the processing area or first POU on the distribution loop 2 Monthly 24 Conductivity Critical Water At the point the distribution loop enters the processing area or first POU on the distribution loop 1 Monthly 12 Total Alkalinity Critical Water After the last treatment step, storage tanks At the point the distribution loop enters the processing area or first POU on the distribution loop 2 Monthly 24 Total Hardness Critical Water After the last treatment step At the point the distribution loop enters the processing area or first POU on the distribution loop 2 Monthly 24 Bacteria Critical Water Loop out point Loop return point Each location of point-of-use in department (subject to change): Decon Sink Scope Sink Unity Ultrasonic Washer-Disinfector #1 Washer-Disinfector #2 7 Monthly 84 Endotoxin Critical Water Water storage tank Loop out point Loop return point Each location of point-of-use in department (subject to change): Decon Sink Scope Sink Unity Ultrasonic Washer-Disinfector #1 Washer-Disinfector #2 8 Monthly 96 Total Organic Carbon (TOC) Critical Water After the last treatment step At the point the distribution loop enters the processing area or first POU on the distribution loop 2 Quarterly 8 Ionic Contaminants Aluminum Chloride Copper Iron Manganese Nitrate Phosphate Sulfate Silicate Zinc Critical Water After the last treatment step At the point the distribution loop enters the processing area or first POU on the distribution loop 2 Annually 2 pH Utility Water At the point the distribution loop enters the processing area or first POU on the distribution loop 1 Quarterly 4 Conductivity Utility Water At the point the distribution loop enters the processing area or first POU on the distribution loop 1 Quarterly 4 Total Alkalinity Utility Water At the point the distribution loop enters the processing area or first POU on the distribution loop 1 Quarterly 4 Total Hardness Utility Water At the point the distribution loop enters the processing area or first POU on the distribution loop 1 Quarterly 4 Bacteria Utility Water Each location of point-of-use in department (subject to change): Decon Sink Scope Sink Unity Ultrasonic Washer-Disinfector #1 Washer-Disinfector #2 5 Quarterly 20 Ionic Contaminants Aluminum Chloride Copper Iron Manganese Nitrate Phosphate Sulfate Silicate Zinc Utility Water At the point the distribution loop enters the processing area or first POU on the distribution loop 1 Annually 2 pH Steam Steam Sterilizer Cooler Sample Port 1 Quarterly 4 Conductivity Steam Steam Sterilizer Cooler Sample Port 1 Quarterly 4 Total Alkalinity Steam Steam Sterilizer Cooler Sample Port 1 Quarterly 4 Total Hardness Steam Steam Sterilizer Cooler Sample Port 1 Quarterly 4 Ionic Contaminants Aluminum Chloride Copper Iron Manganese Nitrate Phosphate Sulfate Silicate Zinc Steam Steam Sterilizer Cooler Sample Port 1 Annually
1 Bacteria Automated Endoscope Reprocessor Medivator #27 Right Basin Medivator #27 Left Basin Medivator #28 Right Basin Medivator #28 Left Basin 4 Monthly 48 Endotoxin Automated Endoscope Reprocessor Medivator #27 Right Basin Medivator #27 Left Basin Medivator #28 Right Basin Medivator #28 Left Basin 4 Monthly 48 MANHATTAN pH Critical Water After the last treatment step at the water generation system At the point the distribution loop enters the processing area or first POU on the distribution loop
2 Monthly 24 Conductivity Critical Water At the point the distribution loop enters the processing area or first POU on the distribution loop 1 Monthly 12 Total Alkalinity Critical Water After the last treatment step at the water generation system At the point the distribution loop enters the processing area or first POU on the distribution loop 2 Monthly 24 Total Hardness Critical Water After the last treatment step at the water generation system At the point the distribution loop enters the processing area or first POU on the distribution loop 2 Monthly 24 Bacteria Critical Water Loop out point Loop return point Each location of point-of-use in department (subject to change): Decon Sink #1 Decon Sink #2 Decon Sink #3 Decon Sink #4 Scope Sink #1 Scope Sink #2 Caviwave PRO Ultrasonic #1 Caviwave PRO Ultrasonic #2 Caviwave PRO Ultrasonic #3 Caviwave PRO Ultrasonic #4 Washer-Disinfector #1 Washer-Disinfector #2 Washer-Disinfector #3 15 Monthly 180 Endotoxin Critical Water Water storage tank Loop out point Loop return point Each location of point-of-use in department (subject to change): Decon Sink #1 Decon Sink #2 Decon Sink #3 Decon Sink #4 Scope Sink #1 Scope Sink #2 Caviwave PRO Ultrasonic #1 Caviwave PRO Ultrasonic #2 Caviwave PRO Ultrasonic #3 Caviwave PRO Ultrasonic #4 Washer-Disinfector #1 Washer-Disinfector #2 Washer-Disinfector #3 16 Monthly 192 Total Organic Carbon (TOC) Critical Water After the last treatment step at the water generation system At the point the distribution loop enters the processing area or first POU on the distribution loop 2 Quarterly 8 Ionic Contaminants Aluminum Chloride Copper Iron Manganese Nitrate Phosphate Sulfate Silicate Zinc Critical Water After the last treatment step at the water generation system At the point the distribution loop enters the processing area or first POU on the distribution loop 2 Annually 2 pH Utility Water At the point the distribution loop enters the processing area or first POU on the distribution loop 1 Quarterly 4 Conductivity Utility Water At the point the distribution loop enters the processing area or first POU on the distribution loop 1 Quarterly 4 Total Alkalinity Utility Water At the point the distribution loop enters the processing area or first POU on the distribution loop 1 Quarterly 4 Total Hardness Utility Water At the point the distribution loop enters the processing area or first POU on the distribution loop 1 Quarterly 4 Bacteria Utility Water Each location of point-of-use in department (subject to change): Decon Sink #1 Decon Sink #2 Decon Sink #3 Decon Sink #4 Scope Sink #1 Scope Sink #2 Caviwave PRO Ultrasonic #1 Caviwave PRO Ultrasonic #2 Caviwave PRO Ultrasonic #3 Caviwave PRO Ultrasonic #4 Washer-Disinfector #1 Washer-Disinfector #2 Washer-Disinvector #3 13 Quarterly 52 Ionic Contaminants Aluminum Chloride Copper Iron Manganese Nitrate Phosphate Sulfate Silicate Zinc Utility Water At the point the distribution loop enters the processing area or first POU on the distribution loop 1 Annually 1 pH Steam Steam Sterilizer Cooler Sample Port 2 Quarterly 8 Conductivity Steam Steam Sterilizer Cooler Sample Port 2 Quarterly 8 Total Alkalinity Steam Steam Sterilizer Cooler Sample Port 2 Quarterly 8 Total Hardness Steam Steam Sterilizer Cooler Sample Port 2 Quarterly 8 Ionic Contaminants Aluminum Chloride Copper Iron Manganese Nitrate Phosphate Sulfate Silicate Zinc Steam Steam Sterilizer Cooler Sample Port 2 Annually 2 Bacteria Automated Endoscope Reprocessor Medivator #1 Right Basin Medivator #1 Left Basin Medivator #2 Right Basin Medivator #2 Left Basin Medivator #3 Right Basin Medivator #3 Left Basin 6 Monthly 72 Endotoxin Automated Endoscope Reprocessor Medivator #1 Right Basin Medivator #1 Left Basin Medivator #2 Right Basin Medivator #2 Left Basin Medivator #3 Right Basin Medivator #3 Left Basin 6 Monthly 72

TABLE 2: Categories and Levels of Water Quality for Medical Device Processing TABLE 3 Critical Water System Performance Qualification Schedule (Intensive Daily)

TABLE 4 Critical Water System Performance Qualification Schedule (Reduced Frequency)
TABLE 5 Frequency for Water Quality Monitoring at Water Generation System
TABLE 6 Frequency For Water Quality Monitoring At Point-Of-Water-Use

Similar Contracts

Same NAICS industry code

NAICS: 221310
New
SLED
Turnkey Exchange Of Spent Granular Activated Carbon (Gac)
Solicitation # P-130000899
The City of San Buenaventura is soliciting proposals for a turnkey exchange of spent Granular Activated Carbon (GAC) in Ventura, California. The contract requires the supplier to provide 40,000 pounds of new 100% Virgin Coconut-Based 12x40 mesh size carbon per change-out, ensuring all materials meet AWWA B604 and ANSI/NSF 61 standards for drinking water. Services must occur at least once and no more than twice per year, with the initial service required within 10 business days of contract execution. The contractor is responsible for the offsite transport of spent GAC using a proper nonhazardous manifest and must maintain OSHA compliant Confined Space Entry Training Certification. Award decisions will be based on the best fit and value, evaluated through contractor capability (40%), reference evaluation (25%), cost of service (25%), and the quality and completeness of the proposal (10%). Proposals must be submitted by November 24, 2026, via both a sealed hard copy and a PDF electronic copy to the Purchasing Supervisor. Payment is processed within 30 days of receipt of an invoice, provided it is accompanied by a post job report including product specification and SDS sheets. The agreement is governed by the City's General Services Agreement, and while bid and performance bonds are not required, contractors must provide a Taxpayer Identification Number and adhere to specific insurance requirements.
City of San Buenaventura

POSTED

about 11 hours ago

DEADLINE

in 26 days
View Details
NAICS: 221310
New
SLED
DESC Well & Water Treatment and Maintenance
Solicitation # RFB DOS 2027-05
The State of New Hampshire Department of Safety, Division of Emergency Services and Communications, is soliciting bids under RFB DOS 2027-05 for well and water treatment operations and management services. The contract covers the maintenance of two wells located at 311 Meredith Road in Laconia, New Hampshire. Required monthly services include physical and comprehensive inspections of the water system from the meter through all filtration equipment, the provision and installation of consumables such as salt and cartridge filters, and the facilitation of water sampling and testing through a state-approved lab according to the NHDES Master Sampling Schedule. Any system deficiencies or concerns must be reported in writing to the State within 24 hours of inspection. The award will be granted to the vendor providing the lowest total cost who meets all terms and conditions. Qualified bidders must hold a New Hampshire Contract Operator Services certification per Administrative Rule Env-Dw 502 and provide documentation of at least two current clients receiving similar services. The contract is governed by New Hampshire law and requires compliance with NH RSA chapter 281-A for workers compensation, as well as comprehensive general liability insurance with minimum coverage of 1 million dollars per occurrence and 2 million dollars aggregate. Bids must be submitted in a searchable PDF format via email by the specified deadline, and vendors must be registered and in good standing with the New Hampshire Secretary of State.
Department Of Safety

POSTED

1 day ago

DEADLINE

in 15 days
View Details
NAICS: 221310
New
Federal
Utilities Privatization of the Water System at U.S. Army Garrison-Hawaii (USAG-HI), Island of Oahu
Solicitation # SP060026R0813
DLA Energy, in coordination with the United States Army, is soliciting proposals for the privatization of the water distribution utility system at U.S. Army Garrison-Hawaii, Island of Oahu. This unrestricted competitive action involves the transfer of ownership, operation, maintenance, and repair responsibilities to a private entity for a maximum term of 50 years. The selected contractor will be responsible for all labor, materials, equipment, and capital investments necessary to ensure continuous and reliable service, including environmental compliance and major system renovations. The process includes a contractor-proposed transition period, after which the system will be conveyed via a Bill of Sale, with real property interests provided as easements or rights of access. Award will be based on a best value determination, where non-price factors—specifically technical capability, past performance, and risk—are significantly more important than price. Technical evaluation focuses on service interruption and contingency planning, operations and maintenance, and system deficiency corrections. Offerors must submit a comprehensive proposal consisting of four volumes, including a technical proposal and a socioeconomic plan demonstrating a commitment to small business subcontracting. Due to the sensitive nature of the requirement, several solicitation documents are secured, requiring vendors to obtain special access through SAM.gov and the Joint Certification Program. Proposals are due by January 7, 2027.
DLA Energy

POSTED

1 day ago

DEADLINE

in 2 months
View Details
NAICS: 221310
New
SLED
TUD Long-Term Water Service ContractThe Tuolumne Utilities District is establishing a long-term water service contract with the United States Bureau of Reclamation to secure Central Valley Project water from the New Melones Reservoir. This agreement is designed to increase water supply reliability and diversify the district's portfolio to support existing service areas and growth envisioned under the Tuolumne County General Plan 2040. The contract allows for the diversion of up to 6,000 acre-feet annually, with 2,500 acre-feet allocated for the district's existing service area and 3,500 acre-feet designated for the Chicken Ranch Rancheria Tribe of Me-Wuk Indians for agricultural, residential, and commercial use within their trust boundaries. The agreement remains effective through December 31, 2051, and utilizes existing pipelines and facilities for water conveyance. Payment is based on a rate structure that includes components for operations and maintenance, storage, water marketing, and dam safety, with tiered pricing applied if annual deliveries exceed 80 percent of the contract total. The district is responsible for constructing necessary conveyance facilities at no cost to the United States and must comply with all federal, state, and local water quality standards. Additionally, the contract mandates strict adherence to civil rights and non-discrimination laws, including the Civil Rights Act of 1964 and the Americans with Disabilities Act.
Tuolumne Utilities District

POSTED

1 day ago

DEADLINE

N/A
View Details
NAICS: 221310
New
SLED
26YS0058 - Supplemental Water Treatment Operator Labor Services
Solicitation # 26YS0058
The California Department of Veterans Affairs (CalVet) is soliciting bids for solicitation 26YS0058 to provide supplemental water treatment operator labor services for the Rector Reservoir Water Treatment Plant in Napa, California. The contract term runs from January 1, 2027, through December 31, 2029. The selected contractor will provide qualified, licensed operators to perform critical plant operations, including daily water treatment testing, dosage determinations, and the maintenance and repair of mechanical equipment such as pumps, chlorinators, and metering devices. Operators must possess a T-2 license from the State Water Resources Control Board and have at least three years of documented experience in a potable water treatment plant. Bids are due by November 5, 2026, at 5:00 PM and must be submitted via email to the designated Contract Analyst. The basis for award will be determined by compliance with all invitation for bid requirements and the verification of all-inclusive hourly rates for regular, overtime, and weekend/holiday time. Required submissions include a mandatory bid rate sheet, bidder certifications, tax IDs, and business licenses. The contractor must be able to respond to service requests within eight hours of notification and ensure all personnel complete a three-day paid orientation. Preference may be given to certified Small or Micro business enterprises, and bidders are encouraged to provide Disabled Veteran Business Enterprise certifications.
California Department of Veterans Affairs

POSTED

4 days ago

DEADLINE

in 7 days
View Details

More opportunities from Department Of Veterans Affairs → 242-NETWORK Contract Office 02 (36C242)

Same awarding agency

NAICS: 236220
New
Federal
Z1DA--Project # 528A6-26-607 Replace Flooring for Rm. 140 B76 | Bath VAMC
Solicitation # 36C24227B0004
The Department of Veterans Affairs, Network Contracting Office 2, is issuing a presolicitation notice for solicitation 36C24227B0004 to replace flooring in Building 76, Rooms 140 and 140A at the Bath VA Medical Center in Bath, New York. The scope of work includes the removal and replacement of existing flooring and electrical modifications to support X-ray equipment, such as extending power conduits and cables, installing new power switches, and providing structural support. The contractor is responsible for providing all supervision, labor, tools, equipment, and materials. Due to the facility being a fully functional medical center, work hours may be restricted, and off-hours work may be required to maintain uninterrupted operations. All work must be completed within 60 days of receiving the Notice to Proceed. This procurement is a 100 percent set-aside for Service-Disabled Veteran Owned Small Businesses (SDVOSB) under NAICS Code 236220, Commercial and Institutional Building Construction, which has a size standard of 45 million dollars. The estimated magnitude of the project is between 25,000 and 100,000 dollars. The solicitation is expected to be posted on SAM.gov on or about October 28, 2026, with an anticipated award date in November 2026. This will be an Invitation for Bid (IFB) requiring bid bonding, and all bids must be submitted electronically. Inquiries should be directed to the Contracting Officer, Idalia Algarin.
Commercial and Institutional Building Construction

POSTED

about 9 hours ago

DEADLINE

in 26 days
View Details
NAICS: 541350
New
Federal
H399--36C242-27-AP-0524| 528-27-1-6893-0009 Elevator Inspections, VA Medical Center (VAMC) & Bath VA Medical Center (VAMC)
Solicitation # 36C24227Q0041
The Department of Veterans Affairs is soliciting quotes for a Qualification of Elevator Inspectors (QEI)-Certified Inspector to serve as a third-party witness for annual and semi-annual safety inspections and testing at the Canandaigua and Bath VA Medical Centers in New York. This service-disabled veteran-owned small business set-aside under NAICS 541350 consists of a 12-month base period starting November 1, 2026, with four additional 12-month option years. The certified inspector is responsible for independent witnessing and reporting using VA Form 10-6167, while a separate maintenance contractor provides the labor and equipment to perform the physical tests. All work must comply with the most current ASME A17.1 and ANSI/ASME 17.2 codes. The scope of work includes semi-annual inspections, annual hydraulic pressure tests, and annual traction no-load tests during the base and final option years, with five-year traction full-load tests scheduled for Bath VAMC in option year one and Canandaigua VAMC in option year two. Services are typically performed in April and October. Award will be based on an evaluation of technical capability, past performance, and total price. Offerors must submit a technical capability statement of no more than 10 pages and adhere to New York State licensing and Service Contract Act wage determinations.
Building Inspection Services

POSTED

about 9 hours ago

DEADLINE

in 4 days
View Details
NAICS: 334516
New
Federal
6550--Sources Sought: Tuberculosis Testing Reagents and Consumables for the Buffalo VA Medical Center. In support of the Tuberculosis (TB) Testing Program
Solicitation # 36C24227Q0095
The Department of Veterans Affairs, VISN 02 Network Contracting Office, has issued a Sources Sought Notice under solicitation number 36C24227Q0095 to identify qualified sources for tuberculosis testing reagents and consumables for the Buffalo VA Medical Center. This request is for information and planning purposes only and does not constitute a formal solicitation. The government is seeking a domestic source capable of providing FDA-cleared reagents, test kits, controls, and processing materials that are compatible with the T-SPOT TB testing platform and methodology. All items must be delivered in original manufacturer packaging with sufficient shelf life and meet all manufacturer specifications for accuracy and performance within an accredited clinical laboratory environment. Interested respondents must submit their capabilities via email to the contracting officer by November 12, 2026. Submissions must include the company's Sam.gov Unique Entity ID, business size status relative to NAICS code 334516, and details regarding whether they are a manufacturer or distributor. The government is specifically looking for information on business representations such as SDVOSB, VOSB, Hubzone, and WOSB status, as well as any existing federal contracts through GSA, NAC, or NASA SEWP. While general pricing is encouraged for market research, it will not be used for award evaluation. Responses must demonstrate clear evidence that any equivalent products meet all salient characteristics and specify if the items are manufactured within the United States.
Analytical Laboratory Instrument Manufacturing

POSTED

about 9 hours ago

DEADLINE

in 14 days
View Details

Ready to Pursue This Opportunity?

Get AI-powered intelligence on this solicitation and the ones like it

Every page of the solicitation package shredded into a compliance breakdown

AI-powered matching based on your capabilities and past performance

Competitor and incumbent history on the requirement

Automated alerts on amendments, Q&A deadlines, and award

Miguel
Hillary
Keith Deutsch
Christine

Join 750+ contractors already using CLEATUS