Full Description
Equipment Requirement: Minimum Required Features and Specifications Type / Intended Use: Must be a transrectal ultrasound transducer specifically designed for prostate imaging and biopsy guidance in urology applications. Must provide multi-plane (triplane) imaging capability allowing visualization of transverse, sagittal, and end-fire planes without changing probes.
Frequency & Imaging Performance: Must operate within a frequency range of approximately 4 MHz to 14 MHz to provide both high-resolution and deep-penetration imaging of the prostate. Must support B-mode, M-mode, Color Doppler, and Power Doppler imaging modes (with optional elastography and contrast imaging if supported by the connected ultrasound system).
Biopsy and Procedural Capability: Must include or support an integrated biopsy guide channel that allows both biplane and end-fire biopsy procedures without requiring probe exchange. Must accommodate standard 17- or 18-gauge biopsy needles through compatible biopsy guide attachments. Must permit real-time needle visualization during transrectal or transperineal prostate biopsy procedures.
4. Design & Ergonomics Must be a compact, endocavity-style probe suitable for patient comfort and operator ergonomics during transrectal use. Must include a control ( Smart ) button or equivalent for hands-free freezing, storing, or switching imaging planes directly from the probe handle. Must be designed for one-handed operation and compatible with sterile probe covers.
5. Compatibility Must be fully compatible with BK Medical ultrasound systems, including bk3000, bk5000, bkActiv, and bkSpecto platforms.
Must connect using the appropriate BK Medical proprietary connector interface without the need for modification or adapters.
6. Cleaning & Sterilization Must be fully immersible and validated for standard high-level disinfection or sterilization processes used for endocavity probes.
Must be compatible with approved sterile covers and disinfection solutions specified in the manufacturer s reprocessing guide.
7. Construction & Safety Must be constructed of biocompatible materials suitable for internal use (per ISO 10993 or equivalent standards). Must comply with applicable IEC/EN 60601-1 electrical safety and ultrasound acoustic output limits. Must be supplied new, manufacturer-certified, and covered under OEM warranty for clinical use.
Summary Table Feature Category Requirement Type Triplane transrectal prostate ultrasound transducer Frequency 4 14 MHz Imaging Modes B, M, Color Doppler, Power Doppler (Elastography optional) Biopsy Capability Integrated dual-plane biopsy guide Needle Compatibility 17 18 G Operation One-handed, Smart Button on probe Sterilization Fully immersible, high-level disinfectable Compatibility bk3000, bk5000, bkActiv, bkSpecto Intended Use Prostate imaging and biopsy guidance Condition New OEM product, OEM warranty
Similar Contracts
Same NAICS industry code
More opportunities from Department Of Veterans Affairs → 248-NETWORK Contract Office 8 (36C248)
Same awarding agency
Ready to Pursue This Opportunity?
Get AI-powered intelligence on this solicitation and the ones like it
Every page of the solicitation package shredded into a compliance breakdown
AI-powered matching based on your capabilities and past performance
Competitor and incumbent history on the requirement
Automated alerts on amendments, Q&A deadlines, and award
Join 750+ contractors already using CLEATUS