K099--Liquid Nitrogen Safety Upgrades

Closed
36C24126Q0086Federal

Contract Overview

Solicitation details, issuing organization, response deadlines, documents, and interested companies for this government contract opportunity.

Active Opportunities Like This One

NAICS: 541330
New
SLED
RFP/RFQ - On-Call Engineering Services for Torrington Transfer Station/NRRA
Solicitation # EDC 11- 2026 -01
The Northwest Hills Council of Governments (NHCOG) is soliciting proposals under solicitation number EDC 11- 2026 -01 for on-call engineering services to support the Torrington Transfer Station and the Northwest Resource Recovery Authority (NRRA). The agency seeks a qualified firm specializing in Municipal Solid Waste and Recycling Transfer Stations to provide technical analysis, facility and operational capacity assessments, site planning, and permitting assistance. These services are intended to support ongoing operations and inform the NRRA Strategic Plan and future capital investment decisions, ensuring compliance with CT DEEP General Permits. Interested firms must submit their proposals by November 30, 2026, at 11:00 AM. Submissions must include a project management plan, key personnel resumes limited to two pages, three relevant references, a technical approach, and a detailed project timeline. A fee proposal containing hourly rates or lump sum allocations must be submitted in a separate sealed envelope. Evaluation will be based on personnel qualifications, experience with multiuse trails and waste facility planning, the proposed project approach, and the fee schedule. Awardees must provide proof of workers compensation, liability, and vehicle insurance, including a waiver of subrogation and naming the State of Connecticut and NHCOG as additional insureds within ten days of the award. Required documentation also includes a Non-Collusion Affidavit and certifications of non-discrimination.
DAS Procurement

POSTED

about 11 hours ago

DEADLINE

in about 1 month

General Info

NAICS

541330 - Engineering Services

Place of Performance

Boston, ME, 02130, USA

Set-Aside

NONE

Documents

1

Sources Sought Notice 36C24126Q0086 Liquid Nitrogen Safety Upgrades

DOCX, Low priority10 pages · sources-sought
Low

AI Contract Breakdown

Uniform Contract Format

Sign up to view the full breakdown with detailed analysis of each section.

Timeline

PhaseClosed
Posted

Sources Sought

Response Deadline

Deadline has passed

Submission Closed

Find active opportunities like this

Start your free trial to discover similar active contracts, track opportunities, and build proposals with AI assistance.

Organization & Contact Information

Show more
AgencyDepartment Of Veterans Affairs → 241-NETWORK Contract Office 01 (36C241)
Contacts1 person available
OfficeTOGUS, ME, 04330, USA
Office AddressTOGUS, ME, 04330, USA
Contacts
Issa ShawkiContracting Officer

Full Description

Show more
This is a Sources Sought notice only. This is not a request for quotes and no contract will be awarded from this announcement. The Government will not provide any reimbursement for responses submitted in response to this Source Sought notice. Respondents will not be notified of the results of the evaluation. If a solicitation is issued it shall be announced at a later date, and all interest parties must response to that solicitation announcement separately. Responses to this notice are not a request to be added to a prospective bidders list or to receive a copy of the solicitation The purpose of this announcement is to perform market research to gain knowledge of potential qualified sources and their size classification relative to NAICS 541330, (K099) with a size standard $25.5 million. The Department of Veterans Affairs (VA), Network Contracting Office 1 (NCO 1) is seeking to identify any vendor capable of providing Liquid Nitrogen Safety Upgrades per the Statement of Work below. This Sources Sought notice provides an opportunity for respondents to submit their capability and availability to provide the requirement described below. Vendors are encouraged to submit information relative to their capabilities to fulfill this requirement, in the form of a statement that addresses the specific requirement identified in this Sources Sought. Information received from this Sources Sought shall be utilized to facilitate the Contracting Officer s review of the market base, for acquisition planning, size determination, and procurement strategy. Submission Instructions: Interested parties who consider themselves qualified to perform the services are invited to submit a response to this Sources Sought Notice by 12PM EST, November 20, 2025. All responses under this Sources Sought Notice must be emailed to contact@example.gov with Sources Sought #36C24126Q0086 in the subject line. Telephone inquiries will not be accepted or acknowledged, and no feedback or evaluations will be provided to companies regarding their submissions. Interested parties should complete the attached Sources Sought Worksheet. Parties may submit additional information related to their capabilities, provided it contains all the requirements contained in the Sources Sought Worksheet. Responses to this Sources Sought shall not exceed 8 pages. In addition, all submissions should be provided electronically in a Microsoft Word or Adobe PDF format. Attachment 1 Sources Sought Worksheet QUALIFICATION INFORMATION: Company / Institute Name: _______________________________________________________ Address: ______________________________________________________________________ Phone Number: ________________________________________________________________ Point of Contact: _______________________________________________________________ E-mail Address: ________________________________________________________________ Unique Entity Identifier (UEI) #: ___________________________________________________ CAGE Code: __________________________________________________________________ SAM Registered: (Y / N) Other available contract vehicles applicable to this sources sought (GSA/FSS/NASA SEWP/ETC): __________________________________________________________________ ______________________________________________________________________________ ______________________________________________________________________________ SOCIO-ECONOMIC STATUS: VIP Verified SDVOSB: (Y / N) VIP Verified VOSB: (Y / N) 8(a): (Y / N) HUBZone: (Y / N) Economically Disadvantaged Women-Owned Small Business: (Y / N) Women-Owned Small Business: (Y / N) Small Business: (Y / N) NOTE: Respondent claiming SDVOSB and VOSB status shall be registered and Center for Veterans Enterprise (CVE) verified in VetBiz Registry www.vetbiz.gov. BASED ON THE RESPONSES TO THIS SOURCES SOUGHT NOTICE/MARKET RESEARCH, THIS REQUIREMENT MAY BE SET-ASIDE FOR SDVOSB, VOSB, SMALL BUSINESSES OR PROCURED THROUGH FULL AND OPEN COMPETITION. STATEMENT OF WORK A. GENERAL INFORMATION 1. Item: Enhancements to the VA Central Biorepository (VACB) liquid nitrogen system to provide a fail-safe environment in case of a liquid nitrogen leak in the Building. The design of the enhanced liquid nitrogen (LN2) system must meet the following criteria: a. Be simple to troubleshoot and easy to reestablish LN2 functionality within a period of less than an hour following a false positive event that led to a LN2 shutdown. b. The system must NOT be complex as to require an expert or extensive amount of time to troubleshoot before functionality can be reestablished. c. System must be designed so that a lab personnel can operate with minimal training. d. System shall contain a fail-safe mechanism in the event of a liquid nitrogen leak. 2. Purpose: The purpose of this contract is to provide certain enhancements necessary to provide a fail-safe environment in the VACB 3. Scope of Work: The contractor shall provide all resources necessary to accomplish the deliverables described in this performance work statement. The contractor should have expertise in cryogenics with positive past performance documented in the CPARS system. The Medical Center requires the services of a qualified contractor which shall provide a fail-safe environment in case of a liquid nitrogen leak in the Building. The VA Central Biorepository (VACB) stores millions of blood samples in a large robotic nitro store, several cryogenic liquid nitrogen tanks and mechanical freezers. The cryogenic tanks and the robotic nitrostore are supplied with liquid nitrogen from a large liquid nitrogen tank (11,000- gallon capacity) that is located outside the building at VABHS. The piping from this external tank to the nitrostore and cryogenic tanks was installed by Maccomber Cryogenics around 2011. Currently there are several oxygen sensors located within the Biorepository that are tied into the alarm systems. When excess liquid nitrogen enters the laboratory the oxygen concentration in the room can drop below levels necessary to sustain life. Currently, the only way to stop the flow of liquid nitrogen is to manually shut off the supply line at the liquid nitrogen tank. If the leak happens during off hours this can take hours leading to an extremely dangerous situation, especially for non-laboratory staff like police or engineering staff that might enter the room to address the alarm. To mitigate risk to staff, under this contract the contractor shall: · Add safety enhancements that would automatically cut-off the liquid nitrogen supply from the main external tank when the oxygen concentration drops below the recommended set-point. · Add 4 E-stop push button stations, · Add three (3) additional oxygen sensors, · Add emergency shutoff valves · Add emergency inlet port for the nitrostore freezer and the cryotank farm room. · All electrical signals emanating from the above mentioned instruments, including an existing Tox Gard ii located in the nitrostore building, shall be connected to the main oxygen monitoring and control panel and be integrated with the Biorepository s Alarm systems as well as to the building s AEM alarm output system. 4. Background: The VACB processes and stores specimens for clinical trials run by the VA Cooperative Studies Program. The Biorepository is involved in a large-scale genomics project that will require processing and storing of up to 1,000 samples a day in addition to existing and ongoing processing for other Cooperative Studies Program projects. The Biorepository stores millions of specimens donated by veterans for clinical research in a large robotic store known as the nitrostore, cryogenic tanks and mechanical freezers. The nitrostore and cryogenic tanks are cooled with liquid nitrogen which is supplied from a large external liquid nitrogen tank located outside the building. Currently there are two shut off valves that are located near the external liquid nitrogen tank that would shut off the supply of liquid nitrogen to the nitrostore building and to the cryogenic tank room. The nitrostore requires a constant supply of liquid nitrogen, so a long shut off of the liquid nitrogen supply to the nitrostore would be detrimental to the stored specimens. We can shut off liquid nitrogen supply to the cryotank room if all the cryotanks are filled with appropriate amounts of liquid nitrogen. There have been several incidences of excess nitrogen releases in the laboratory where the oxygen concentration in the VACB dropped below the set-point of the oxygen sensors. This was due to both equipment malfunction and filling of the liquid nitrogen tank. All events occurred during off hours. The drop in oxygen concentration is not compatible with human life. While biorepository staff are well trained to handle this event, first responders such as police or engineering staff who respond to the low oxygen alarm may enter the Biorepository without realizing the consequences. 5. Performance Period: Four months. No options. 6. Type of Contract: Firm-fixed price Service contract 7. Place of Performance: Service performed at VA Boston Healthcare System, 150 S. Huntington Avenue, Jamaica Plain, MA 02130. Room RB-8 8. General Requirements: After several discussions the Biorepository and the local engineering service decided to go with pneumatic valves. The pressure required to operate these pneumatically controlled valves will be enabled with two nitrogen gas cylinders which will be placed near the liquid nitrogen tank. These nitrogen gas cylinders should be protected from external elements and secured properly by the vendor. In addition, the vendor shall provide an emergency inlet port that will supply liquid nitrogen to the nitrostore and the cryotank farm when the system is down for maintenance. In addition, we would require the vendor to provide all engineering services, documentation deliverables, instrument installation and configuration, required electrical and mechanical installation, project management at site and internally, validation of the system, necessary training and a one year system warranty that includes parts and labor for the system. The figure below shows how the two valves are controlled by the two N2 cylinders. We will not use LN2 supply pressure to activate these valves.
9. Requested Enhancements Control System Scope of Supply a. One (1) Main O2 Monitoring & LN2 Isolation Control Panel b. Two (2) Local O2 Monitoring Control Panels [for redundant O2 monitors] c. Three (3) MSA TG5000 O2 Monitors with remote sensor kits d. Three (3) MSA Toxguard II remote sensor kits e. Four (4) NEMA 4 Lighted E-Stop Pushbutton Stations f. Two (2) LN2 Electronic Pressure Transmitters with manifold ranged 0 - 60 PSIG g. One (1) N2 Electronic Pressure Transmitter with manifold ranged 0 - 3,000 PSIG h. One (1) custom engineered LN2 valve manifold with the following features: o All piping vacuum insulated for LN2 service o Dual LN2 isolation valves, spring fail-closed pneumatic operator, 24VDC solenoids, DPDT Open & Close Limit switches and manual override capability o Pressure relief valves in all pipe sections downstream of valves o Emergency supply ports for the Nitrostore and Cryotank LN2 supply piping o Taps for LN2 tank supply and building LN2 delivery pressure transmitters i. N2 tank control and monitoring system with the following features: o Parker Gas Change-Over System unit that allows depressurizing and swapping tanks without loss of N2 pressure/flow. When one is empty the unit will automatically start drawing from the second tank. o Low pressure switches for each tank s inlet connection o Outdoor weather-shielded lockable cage with room for six (6) tanks total o Two (2) 300CF series replaceable tanks of N2 Engineering Services j. Design and Drafting of Local O2 Monitoring Panel k. Design and Drafting of Main O2 Monitoring & LN2 Isolation Panel l. Programming of PLC controller and HMI Monitoring Screens m. Design and drafting of custom LN2 valve manifold and associated site piping n. Design and drafting of N2 tank monitoring and control system o. Design coordination with AEM and Rees Scientific p. Factory Acceptance Test (FAT) q. Site commissioning, testing and system validation Documentation Deliverables r. Control panel drawings and equipment data sheets s. Control system overview drawings t. LN2 valve manifold piping and instrumentation drawing (P&ID) u. N2 Tank control and monitoring system v. O2 monitoring and LN2 isolation system validation test plan Instrumentation Installation and Configuration w. Install and configure the following equipment and instruments: o MSA TG5000 O2 monitors and remote sensor kits o Remote O2 sensor kits for existing Toxguard IIs o LN2 supply and delivery pressure transmitters o N2 gas tank A pressure transmitter o N2 gas tank pressure switches x. Configure the existing MSA Toxguard IIs and rewire the 115VAC power to be from the UPS in the Main O2 Monitoring and LN2 Isolation Panel. Electrical Installation y. Installation of NECI local and main control panels, remote E-stop stations, and O2 monitors z. Installation of O2 sensor extension kits on existing Toxguard II O2 monitors aa. Supply, installation and wiring termination of all control cables, and power cables shown on the System Overview drawing. bb. Power and signal wiring to the AEP and REES Scientific panels. cc. Supply, installation and wiring termination of control cables to existing alarm sirens and strobe lights Mechanical Installation dd. Installation of LN2 pressure transmitters and N2 pressure transmitter, transmitter sunshade and associated impulse tubing. ee. Installation of the custom outdoor LN2 valve and instrumentation manifold assembly ff. Moving of the Cryotank Room LN2 piping take-off to be downstream of the isolation valves. gg. Addition of on-pad and off-pad pipe supports including foundations as required. hh. Supply and install N2 tank enclosure, N2 gas tank Changeover System and associated piping. Internal and On-Site Project Management ii. Internal project management for detailed system design, parts procurement and regular updates to the VA. jj. Management of electrical, LN2 piping and third-party suppliers (LN2 Manifold Supplier, AEM, Rees Scientific) kk. One-site construction and installation management System Warranty ll. Include a one-year parts and labor warranty for the system.

More opportunities from Department Of Veterans Affairs → 241-NETWORK Contract Office 01 (36C241)

Same awarding agency

NAICS: 532490
New
Federal
W038--Wheeled Rental Loader - Winter FY27
Solicitation # 36C24126Q0815
The VA Maine Healthcare System Togus is seeking a qualified vendor under a total set-aside for Service-Disabled Veteran-Owned Small Businesses (SDVOSB) to provide month-to-month rental and support services for two wheeled loaders and snow removal attachments. The equipment, specified as CAT 926 or CAT 930 models weighing between 27,972 and 31,124 lbs, will be used for snow removal at the Togus VAMC campus in Augusta, Maine. Requirements include a 3-yard bucket with a Fusion coupler, 160 to 170 hp Tier 4 Final engines, ROPS/FOPS certified cabs, and liquid-filled tires. The contractor must provide 24/7 service and support, with repairs initiated within 24 hours of notification or the provision of a replacement loader at no additional cost. All preventive maintenance, repairs, and delivery/pickup charges must be included in the firm fixed price quote. The procurement is governed by NAICS code 532490 and follows FAR subpart 12.6 streamlined procedures. Award will be based on a comparative evaluation of technical capability, past performance, and price, with the contract awarded to the responsible vendor who is technically acceptable and offers the lowest evaluated price. Vendors must submit a firm fixed price quote, a technical capability statement, and a signed Limitation of Subcontracting certificate. Strict security and health protocols apply, including mandatory security badging, adherence to VA non-smoking policies, and proof of vaccinations for influenza, COVID-19, and tuberculosis testing for all personnel accessing the station. All quotes must be submitted to the designated contact by November 2, 2026, at 5:00 PM EST.
Other Commercial and Industrial Machinery and Equipment Rental and Leasing

POSTED

about 16 hours ago

DEADLINE

in 6 days
View Details
NAICS: 561210
New
Federal
F109--Water Management Program and Chemical Testing Amendment 001- Q&A
Solicitation # 36C24127Q0016
The Department of Veterans Affairs Network Contracting Office 01 is soliciting quotes for a new Water Management Program and Chemical Testing contract to support three VAMC Boston Healthcare System campuses in Jamaica Plain, West Roxbury, and Brockton. This procurement is a 100% set-aside for Service-Disabled Veteran-Owned Small Businesses (SDVOSB) and Veteran-Owned Small Businesses (VOSB) under NAICS code 561210. The contractor will be responsible for developing, testing, and maintaining a comprehensive Water Management Program, including monthly and quarterly water testing and the implementation of a web-based data management system. All services and reporting must align with AAMI ST-108:2023 standards. The anticipated period of performance consists of a base year from March 1, 2027, to February 29, 2028, with four one-year option renewals. Amendment 0001 updated the Statement of Work, added a Questions and Answers document, and extended the response deadline to November 6, 2026, at 12:00 PM EST. Award will be based on a comparative evaluation of technical quality, qualifications, past performance, veterans involvement, and price. Key technical requirements include the use of an accredited laboratory located within the contiguous United States (east of or within MN, IA, MO, AR, and LA), the assignment of dedicated on-site personnel for monthly testing, and a demonstration of independence with no conflict of interest regarding the sale of treatment chemicals. Prospective offerors must be registered in SAM and the SBA VetCert database and must complete VA vendorization via Form 10091 to be eligible for award.
Facilities Support Services

POSTED

about 16 hours ago

DEADLINE

in 10 days
View Details

Earlier notices in this NAICS code

Posted before this notice, open and closed

Find Active Opportunities Like This

Get AI-powered intelligence on the opportunities still open

Every page of the solicitation package shredded into a compliance breakdown

AI-powered matching based on your capabilities and past performance

Competitor and incumbent history on the requirement

Automated alerts on amendments, Q&A deadlines, and award

Miguel
Hillary
Keith Deutsch
Christine

Join 750+ contractors already using CLEATUS