Q999--36C255-21-AP-2386 - 589-21-3-1424-0040 - KC - Audioscan REM/HIT

Awarded
36C25521Q0365Federal

Contract Overview

Solicitation details, issuing organization, response deadlines, documents, and interested companies for this government contract opportunity.

Active Opportunities Like This One

General Info

Contract Value

$86,880

NAICS

423450 - Medical, Dental, and Hospital Equipment and Supplies Merchant Wholesalers

Place of Performance

KS

Set-Aside

NONE

Award Issued Date

Documents

1

36C25521Q0365.docx

DOCX

AI Contract Breakdown

Uniform Contract Format

Contract not broken down yet

CLEATUS splits the solicitation into Uniform Contract Format (UCF) sections: scope, pricing, deliverables, and evaluation criteria.

Timeline

PhaseAwarded
Posted

Presolicitation

Type Changed

Presolicitation → Solicitation

Awarded

Contract was awarded

Submission Closed

Find active opportunities like this

Get full access to discover similar active contracts, track opportunities, and build proposals with AI assistance.

Organization & Contact Information

Show more
AgencyDepartment Of Veterans Affairs → 255-NETWORK Contract Office 15 (36C255)
Contacts1 person available
OfficeLEAVENWORTH, KS, 66048, USA
Office AddressLEAVENWORTH, KS, 66048, USA
Contacts
Jennifer A SotomayorContract Specialist

Full Description

Show more
1 Kansas City VA Medical Center 4801 E. Linwood Blvd. Kansas City, MO 64128 Audiology SCOPE OF WORK AND GENERAL SPECIFICATIONS FOR: HEARING INSTRUSTMENT VERIFICATION SYSTEM 1. Background The Lenexa CBOC is opening in June 2021 and Audiology is asking for two hearing instrument verification systems to be placed in each hearing aid fitting room. 2. Specifications DESCRIPTION: HEARING INSTRUMENT FITTING SYSTEM MUST MEET OR EXCEED SALIENT CHARACTERISTICS BELOW: REQUIRED FEATURES: Dual probes for fast binaural fitting Viewport featuring Speechmap, patented directional test, unique noise reduction and feedback suppression tests. Test signal for frequency manipulating instruments Fitting protocols (DSL 5.0, NAL-NL1, Camfit) True two channel measurements Single/Dual view display options External monitor capability Sensory Loss Simulator ANSI, ITE & BTE couplers Occlusion effect test Battery Stimulator RECD transducer Probe monitor Telecoil test Manual Gain, output, and distortion tests DIMENSION REQUIREMENTS: Overall dimensions Rear main unit: 14.5 x 16 x 6.5in Test Chamber: 14 x 14.4 x 4in Weight: 17.6 lbs UTILITY REQUIREMENTS: Power consumption: 100-240V, 50-60Hz, 250VA Temperature: -20 to +60C Relative humidity (non-condensing): 5 95% GENERAL REQUIREMENTS: Display type: fluorescent backlit active color Display size: 12.1 diagonal Resolution: 800 x 600 (SVGA) Internal Printer: 3 thermal line printer, 200 dots/inch Headphone amplifier : 2 @ 5 watts each Stimulus channels: 2 Measurement channels: 2 Connectors: 1 ethernet (RJ45) 4 USB 1 HDMI 2 external speakers 2 expansion jacks (1/4 mono) 2 real ear mics (3.5mm st) 1 RECD transducer (3.5mm st) 1 monitor headset (3.5mm st) 1 test chamber ref. mic (3.5mm st) 1 coupler mic (3.5mm st) 1 battery substitute (3.5mm st) TEST BOX REQUIREMENTS Working space: 11 x 5 x 1.5in 2 independent speakers 2 x 3in Test Box isolation @1K Hz: .25dB Induction Coils per ANSI S3.22-2003: 1 telephone magnet field sim (TMFS) & 1 23cm diameter test loop Battery Simulator: per ANSI 3.22-2003 Frequency Range: 200-8000Hz Test Stimuli: Tone, Tone Burst, Pink Noise, Dual Direction Noise, Calibrated or Live Speech, ISTS Test Stimulus Levels: 40 to 90dB in 5dB steps Test Stimulus Levels (inductive): 31.6mA/m per ANSI S3.22 - 2003 Test Stimulus Distortion: <2% at 90B SPL and 0.5% at 70dB SPL Test Stimulus Accuracy at Reference Mic for tones (200-2000Hz)+1.5dB SPL Test Stimulus Accuracy at Reference Mic for tones (2000-8000Hz)+2.5dB SPL Equalization Method: Real time modified pressure method (stored for open fittings) Analysis frequencies per octave: 12 Analysis filter bandwidth (noise): 1/12 octave Measurement accuracy at 1kHz for tones: +/-1dB Measurement accuracy re 1kHz for tones: +/-1dB (200-5000Hz) and +/-2.5dB (5000-8000Hz) Measurement Range: 30-140dB SPL Harmonic Distortion measurement: 2nd and 3rd or 2nd plus 3rd Harmonic distortion range: 200-4000Hz Harmonic distortion accuracy: +/-1% Battery Drain range: 0-20 mA Battery Drain accuracy: +/-5% Battery Drain resolution: +/-.01 mA ANSI S3.22 1996 and 2003 tests available: OSPL90 Full-on Gain Reference Test Gain Frequency Response Frequency Range Maximum OSPL90 Harmonic Distortion Attack & Release Time Equiv. Input Noise Input/Output Curves Coupler SPL-Telephone Simulator Simulated Telecoil Sensitivity Battery Drain OTHER Tests Available: Speechmap* real ear audibility measures Real-Time adaptive directional verification Coupler SPL vs. frequency Coupler gain vs. frequency Spectral Analysis Noise reduction verification Distortion vs frequency Manual measurement of output, gain, and distortion VIEWPORT audibility, directional, noise reduction, and feedback suppression verification. ON-EAR MEASURES Speakers: 2 2x3 in. (5x7.5cm) ducted ports Probe microphone tube: Silicone 1mm diameter x 75mm Probe modules: 2 each containing probe and reference microphones Probe microphone noise floor (200-8000Hz): <45dB SPL Frequency range: 200-8000Hz Test Stimuli: frequency modulated tone, tone burst, pink noise, dual directional noise, calibrated speech, live speech, ISTS Frequency modulation: sawtooth +/-3% over 128ms Test stimulus levels at reference microphone for tones at 1m: 40-90dB SPL in 5 dB steps Test stimulus accuracy at reference microphone for tones (200-2000Hz): +/-1.5dB SPL Test stimulus accuracy at reference microphone for tones (2000-8000Hz): +/-2.5dB SPL Equalization method: real time modified pressure method (stored for open fittings) Analysis frequencies per octave: 12 Frequencies per octave (tone burst): 3 Analysis filter bandwidth (speech, noise): 1/3 octave Measurement accuracy at 1kHz for tones: +/-1dB Measurement accuracy at 1kHz for tones: +/-1dB (200-5000Hz) Measurement accuracy at 1kHz for tones: +/-dB (5000-8000Hz) Measurement range: 20-135dB SPL (200-2500Hz) & 30-140dB SPL (2500-8000Hz) ANSI S3.46 1997 TESTS AVAILABLE Real Ear unaided Response Real Ear Aided Response Real Ear Occluded Response Real Ear Insertion Gain OTHER TESTS AVAILABLE Speechmap* real ear audibility measures Real-Time adaptive directional verification Coupler SPL vs. frequency Coupler gain vs. frequency Spectral Analysis Noise reduction verification Distortion vs frequency Manual measurement of output, gain, and distortion VIEWPORT audibility, directional, noise reduction, and feedback suppression verification. FITTING METHODS AVAILABLE Speechmap with DSL 5.0a, NAL-NL1, CAMFIT, Insertion gain with NAL-RP, NAL-NL1, Fig 6, PogoII, Berger, Libby 3. DELIVERY & LOCATION INFORMATION Kansas City VA - Lenexa CBOC Kansas City VA Medical Center Audiology Clinic (Permanent Location) Audiology Clinic (Shipping Location) 15512 W. 113th St. 4801 E. Linwood Blvd. Audiology Lenexa, KS 66219 Kansas City, MO 64128

More opportunities from Department Of Veterans Affairs → 255-NETWORK Contract Office 15 (36C255)

Same awarding agency

NAICS: 236220
New
Federal
Z2DA--657-27-103JB, FCA Replace Building 51 Roof, JB
Solicitation # 36C25527Q0044
The Department of Veterans Affairs, Network Contracting Office 15, is conducting market research through a Sources Sought Notice to determine the availability of qualified Service-Disabled Veteran-Owned Small Business (SDVOSB) firms for a construction project at the St. Louis Medical Center Jefferson Barracks. The project involves the replacement of the roof and associated components for Building 51, covering approximately 56,000 square feet. The scope includes installing a 70-mil thermoplastic polyolefin (TPO) roofing system, replacing four skylights, and updating the penthouse roof-access ladder to meet OSHA and VA safety standards. The estimated value of the resulting contract is between 2 million and 5 million dollars, with a required performance period of 222 calendar days from the Notice to Proceed. The project requires the contractor to manage work within an occupied 24/7 facility, necessitating a detailed phasing plan and strict adherence to infection control and HVAC-intake protection. Key technical requirements include providing a 20-year no-dollar-limit warranty, ensuring positive drainage to prevent standing water for more than 48 hours, and submitting as-built record drawings in searchable PDF and editable CAD formats. Interested firms must be registered in SAM.gov under NAICS code 236220 and provide a capability statement by November 9, 2026, including evidence of CVE certification, bonding capability, and a history of comparable work performed within the last five years.
Commercial and Institutional Building Construction

POSTED

about 14 hours ago

DEADLINE

in 11 days
View Details

Earlier notices in this NAICS code

Posted before this notice, open and closed

Find Active Opportunities Like This

Get AI-powered intelligence on the opportunities still open

Every page of the solicitation package shredded into a compliance breakdown

AI-powered matching based on your capabilities and past performance

Competitor and incumbent history on the requirement

Automated alerts on amendments, Q&A deadlines, and award

Miguel
Hillary
Keith Deutsch
Christine

Join 750+ contractors already using CLEATUS